Operator's Manual

Table Of Contents
3 - 12 BeneVision N1 Patient Monitor Operator’s Manual
1. Select the Main Menu quick key from the Display column select Exit Night Mode.
2. Select Ok.
NOTE
If your monitor is connected to the CMS, it automatically exits the night mode when being
disconnected from the CMS.
The monitor resumes the previous settings after exiting the night mode.
3.7.5 Standby Mode
You can temperately stops patient monitoring without switching off the monitor by entering the standby mode.
3.7.5.1 Entering the Standby Mode
1. Select the Standby quick key, or select the Main Menu quick key from the Patient Management
column select Standby.
2. Define where the patient is by selecting a location in the drop down list when the monitor enters the
standby mode.
3. Select Ok.
The monitor behaves as follows after entering the standby mode:
Stops all parameter measurements.
Disables all the alarms and prompt messages, except for the battery low alarm.
Turns screen brightness to the dimmest after entering the standby mode for 30 seconds.
WARNING
Pay attention to the potential risk of placing the monitor to standby. In the standby mode, the
monitor stops all parameter measurements and disable all the alarm indications, except for the
battery low alarm.
3.7.5.2 Changing the Patient Location at Standby
If you need to change the patient’s location, select patient location from the standby screen.
3.7.5.3 Exiting the Standby Mode
To exit the standby mode, choose any of the following ways:
Select Resume Monitor to exit the standby mode and resume monitoring the current patient.
Select Discharge Patient to discharge the current patient.
3.7.6 Outdoor Mode
The outdoor mode is intended for transferring patients outdoors. The monitor behaves as follows after entering
the outdoor mode:
The parameter color is white and unchangeable.
The screen brightness is automatically changed to 10.
3.7.6.1 Entering the Outdoor Mode
To enter the outdoor mode, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Other tab.