Operator's Manual

Table Of Contents
F - 9
F.3 The Accuracy of Nellcor SpO
2
Sensors
Modified Bland-Altman for SpO
2
- MAX-A, MAX-P Sensors (No Motion): SaO
2
vs. (SpO
2
-
SaO
2
)
Sensor Motion # of Data Points A
RMS
(%) SpO
2
Acceptance
Criteria 70% - 100%
(%)
Pass/Fail
MAX-A, MAX-P No 976 1.54 ≤2.0 Pass
MAX-I, MAX-N No 723 1.41 ≤2.0 Pass
DS-100A No 411 2.16 ≤3.0 Pass
D-YS, OXI-P/I, OXI-A/N No 458 1.96 ≤3.5 Pass
1. SpO
2
- SaO
2
(%) 2. SaO
2 (%)
3. Upper 95% LoA 4. Mean Bias 5. Lower 95% LoA