D - 8 BeneVision N1 Patient Monitor Operator’s Manual
Note: XX represents 2.5 V, 3.3 V,5 V, or 12 V.
D.2.10 Printer Technical Alarm Messages
D.2.11 Technical Alarm Messages Related to Networked Monitoring
Note: XX refers to the department name, YY refers to the room number, and ZZ refers to the bed number.
D.2.12 Other System Technical Alarm Messages
XX V Too High High C There is a problem with the system power supply.
Restart the monitor.
XX V Too Low High C
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Printer Buffer Full Prompt / The printer buffer is full. Wait till the printer
finishes the printing task.
Fail Prompt / The printer runs out of paper or cannot be
connected. Check the printer.
Printing Stopped Prompt / Printing is manually stopped.
Printer Unavailable Prompt / The printer may fail. Check the printer.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
No CMS Low B The monitor is disconnected from the CMS.
Check the network connection.
View Bed XX YY-ZZ,
Network Disconnected.
Low A The network is interrupted when the monitor is
viewing the remote device. Check the network
connection.
Viewed by Bed XX YY-ZZ,
Network Disconnected.
Low A The network is interrupted when the monitor is
viewed by another remote device. Check the
network connection.
WLAN IP Address Conflict Low C Wireless network IP network conflicts. Check the
network settings.
LAN1 IP Address Conflict Low C Wired network LAN1 IP network conflicts. Check
the network settings.
Fail To Get WLAN IP
Address
Low C Unable to automatically obtain the wireless
network IP address. Check the network settings.
Fail To Get LAN1 IP
Address
Low C Unable to automatically obtain the wired
network LAN1 IP address. Check the network
settings.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Storage Card Error High C The storage card fails or files are damaged.
Restart the monitor to format the storage card. If
the alarm persists, contact your service
personnel.