C - 12 BeneVision N1 Patient Monitor Operator’s Manual
C.7.2 Sidestream CO
2
Default Settings
C.7.3 Microstream CO
2
Default Settings
C.7.4 Mainstream CO
2
Default Settings
Speed 6.25 mm/s
Scale 50 mmHg
Waveform Type Draw
Item Default Setting
Item Default Setting
EtO
2
Alarm switch On
High limit 88%
Low limit 18%
Priority Med
Alarm Outputs Off
FiO
2
Alarm switch On
High limit Adult and pediatric: 100%
Neonate: 90%
Low limit 18%
Priority Med
Alarm Outputs Off
BTPS Compensation Off
O
2
Compensation OR: 100%
Other departments: 21%
N
2
O Compenation 0%
AG Compenation 0%
Auto Standby 60 min
Operating Mode Measure
Item Default Setting
BTPS Compen Off
Maximum Hold 20 sec
Auto Standby Off
Operating Mode Measure
Item Default Setting
Maximum Hold 10 sec
O
2
Compensation Off