BeneVision N1 Patient Monitor Operator’s Manual C - 11
C.7 CO
2
Default Settings
C.7.1 General Settings
Scale(mmHg) CVP/pCVP/ICP/RAP/
LAP/UVP venous
pressure
0-20
Art/pArt/Ao/BAP/FAP/
LV/P1/P2 arterial
pressure
0-160
UAP/P3/P4 venous
pressure
0-80
PA 0 -30
PPV Measure Off
PPV Source Auto
PAWP Reference Waveform 1 II
Reference Waveform 2 Resp
Speed 12.5 mm/sec
PA Scale(mmHg) 0-30
Overlapping
Waveform Setup
Left Scale(mmHg) 0-160
Right Scale(mmHg) 0-20
CVP Scale(mmHg) 0-20
ICP Scale(mmHg) 0-20
PA Scale(mmHg) 0-30
Speed 25 mm/sec
Gridlines Off
Item Default Setting
Item Default Setting
EtCO
2
Alarm switch On
High limit Adult and pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult and pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO
2
Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
Zero RR Delay Adult and pediatric: 20 s
Neonate: 15 s
RR Source Auto