Operator's Manual
Table Of Contents
- 1 Safety
- 2 Equipment Introduction
- 2.1 Indications for Use
- 2.2 Applied Parts
- 2.3 Main Unit
- 2.4 Modular Rack
- 2.5 Dock
- 2.6 External Parameter Modules
- 2.7 Installlation
- 2.7.1 Installing the N1 or External Parameter Module into the Modular Rack
- 2.7.2 Removing the N1 or External Parameter Module from the Modular Rack
- 2.7.3 Installing the Modular Rack to the Dock
- 2.7.4 Removing the Modular Rack from the Dock
- 2.7.5 Installing the N1 to the Dock
- 2.7.6 Removing the N1 from the Dock
- 2.8 N1 in Use with the Host Monitor
- 2.9 N1 in Use with the Transport Dock
- 2.10 Input Devices
- 2.11 Printing Devices
- 3 Getting Started
- 3.1 Equipment Preparation Safety Information
- 3.2 Unpacking and Checking
- 3.3 Environmental Requirements
- 3.4 Setting Up the Equipment
- 3.5 Operation and Navigation
- 3.5.1 Using the Touchscreen
- 3.5.2 Using the Mouse
- 3.5.3 Using the On-Screen Keyboard
- 3.5.4 Using the Barcode Reader
- 3.5.4.1 Clearing Old Data Formats (for the Mindray Custom 2D Barcode Reader)
- 3.5.4.2 Establish the Relationship between Patient Demographics and 2D Barcode Fields (for the Mindray Custom 2D Barcode Reader)
- 3.5.4.3 Setting the Barcode Reader Information
- 3.5.4.4 Identifying the Barcode Reader (for the non-Mindray Custom 2D Barcode Reader)
- 3.5.4.5 Selecting Patient Information output by the Barcode Reader (for the Mindray Custom 2D Barcode Reader)
- 3.6 Screen Display
- 3.7 Operating Modes
- 3.8 Configuring Your Monitor
- 3.9 Starting Monitoring a Patient
- 3.10 Stopping a Parameter Measurement
- 3.11 General Operation
- 3.12 Using the On-Screen Timers
- 3.13 Using the External Display
- 3.14 The OxyCRG Screen (only available for the independent external display)
- 3.15 Turning Off the Monitor
- 4 Managing Patients
- 4.1 Discharging a Patient
- 4.2 Admitting a Patient
- 4.3 Managing Patient Information
- 4.3.1 Entering the Patient Management Menu
- 4.3.2 Editing Patient Information
- 4.3.3 Loading Patient Information from the CMS
- 4.3.4 Loading Patient Information from the ADT Server
- 4.3.5 Selecting Query Criteria for Searching Patients in ADT Server
- 4.3.6 Changing Patient Management Settings
- 4.3.7 Setting Monitor Information
- 4.3.8 Setting Monitor Location
- 4.3.9 Setting Patient Location
- 4.3.10 Hiding Patient Name
- 4.4 Transferring Patient Data
- 4.5 Setting the Dock
- 4.6 Connecting the CMS
- 5 Managing Configurations
- 6 Networked Monitoring
- 6.1 Network Introduction
- 6.2 Network Safety Information
- 6.3 Connecting the Monitor to the CMS
- 6.4 Connecting the eGateway
- 6.5 Using the ADT Gateway
- 6.6 Sending Realtime Data, Waveforms and Alarms via HL7 Protocol
- 6.7 Viewing Other Patients (Only available for the independent external display)
- 6.8 Configuring the Network
- 6.8.1 Setting the Wireless Network
- 6.8.2 Disconnecting the Wireless Network
- 6.8.3 Selecting WLAN Band and Channels
- 6.8.4 Managing Certifications
- 6.8.5 Setting Multicast Parameters
- 6.8.6 Setting the Network Service Quality Level
- 6.8.7 Selecting Data Encryption Type
- 6.8.8 Disabling the Broadcasting Patient Demographics Function
- 6.9 MLDAP
- 7 Alarms
- 7.1 Alarm Introduction
- 7.2 Alarm Safety Information
- 7.3 Understanding the Alarms
- 7.4 Accessing On-screen Help for Technical Alarms (AlarmSight)
- 7.5 Checking Physiological Alarm List
- 7.6 Changing Alarm Settings
- 7.6.1 Setting Parameter Alarm Properties
- 7.6.2 Setting Alarm Tone Properties
- 7.6.3 Initiating Auto Alarm Limits
- 7.6.4 Setting the Alarm Delay Time
- 7.6.5 Adjusting the Alarm Light Brightness
- 7.6.6 Setting the Length of Printing Duration On Alarm
- 7.6.7 Setting the Switch of the CMS and eGateway Disconnection Alarm
- 7.6.8 Setting the Priority of the CMS and eGateway Disconnection Alarm
- 7.6.9 Setting the Switch of the SpO2 Desat Alarm Off
- 7.6.10 Setting the Switch of the Zero RR Alarm Off
- 7.7 Pausing Alarms/Pausing Alarm Tones
- 7.8 Resetting Alarms
- 7.9 Latching Alarms
- 7.10 CPB Mode
- 7.11 Intubation Mode
- 7.12 Managing Alarms from Remote Devices (only available for the independent external display)
- 7.12.1 Setting the Tone Pattern for Alarms from Remote Devices
- 7.12.2 Selecting the Alarm Reminder for Remote Devices
- 7.12.3 Presenting Alarm Sound for Remote Devices as per Alarm Priority
- 7.12.4 Resetting Alarms for Remote Devices
- 7.12.5 Authorizing the Alarm Reset to Other Devices
- 7.12.6 Switching Off the Remote Device Disconnection Alarm
- 7.13 Testing Alarms
- 7.14 Actions When an Alarm Occurs
- 8 Monitoring ECG, Arrhythmia, ST and QT
- 8.1 ECG Introduction
- 8.2 ECG Safety Information
- 8.3 ECG Display
- 8.4 Preparing for ECG Monitoring
- 8.5 Changing ECG Settings
- 8.5.1 Choosing an ECG Screen
- 8.5.2 Setting ECG Alarm Properties
- 8.5.3 Changing ECG Wave Settings
- 8.5.4 Disabling the Smart Lead Off Function
- 8.5.5 Setting the Priority of the ECG Lead Off Alarm
- 8.5.6 Adjusting the QRS Volume
- 8.5.7 Adjusting the Minimum QRS Detection Threshold
- 8.5.8 Selecting the ECG Standard
- 8.6 Monitoring Arrhythmia
- 8.7 ST Segment Monitoring
- 8.8 Changing ST Settings
- 8.9 QT/QTc Interval Monitoring
- 8.10 ECG Relearning
- 8.11 Calibrating ECG
- 8.12 Defibrillation Synchronization Pulse Output
- 8.13 ECG Troubleshooting
- 9 Resting 12-Lead ECG Analysis
- 9.1 Resting 12-Lead ECG Analysis Introduction
- 9.2 Entering the 12-Lead Screen
- 9.3 Initiating Resting 12-Lead ECG Analysis
- 9.4 Changing 12-Lead ECG Analysis Settings
- 9.5 Glasgow Resting 12-lead ECG Analysis Algorithm Settings
- 9.6 Saving the 12-Lead Interpretation Report
- 9.7 Printing the 12-Lead Interpretation Report
- 9.8 Exiting the ECG 12-Lead Screen
- 10 Monitoring Respiration (Resp)
- 11 Monitoring Pulse Oxygen Saturation (SpO2)
- 11.1 SpO2 Introduction
- 11.2 SpO2 Safety Information
- 11.3 SpO2 Measurement Limitations
- 11.4 SpO2 Display
- 11.5 Preparing for SpO2 Monitoring
- 11.6 Changing the SpO2 Settings
- 11.6.1 Changing the SpO2 Alarm Settings
- 11.6.2 Nellcor Sat-Seconds Alarm Management
- 11.6.3 Setting the Nellcor SpO2 Sat-Seconds
- 11.6.4 Setting SpO2 Sensitivity (for Masimo SpO2 Module)
- 11.6.5 Enabling FastSAT (for Masimo SpO2)
- 11.6.6 Displaying SIQ (for Masimo SpO2)
- 11.6.7 Changing the Sensitivity
- 11.6.8 Monitoring SpO2 and NIBP Simultaneously
- 11.6.9 Changing the Sweep Speed of the Pleth Wave
- 11.6.10 Setting the Alarm Priority for SpO2 Sensor Off Alarm
- 11.6.11 Setting the SpO2 Tone Mode
- 11.7 Changing the PR Settings
- 11.8 SpO2 Troubleshooting
- 11.9 Nellcor Information
- 12 Monitoring Temperature (Temp)
- 13 Monitoring Noninvasive Blood Pressure (NIBP)
- 13.1 NIBP Introduction
- 13.2 NIBP Safety Information
- 13.3 NIBP Measurement Limitations
- 13.4 Measurement Modes
- 13.5 NIBP Display
- 13.6 Preparing for NIBP Measurements
- 13.7 Starting and Stopping NIBP Measurements
- 13.8 Changing NIBP Settings
- 13.8.1 Setting the NIBP Alarm Properties
- 13.8.2 Setting the Initial Cuff Inflation Pressure
- 13.8.3 Setting the NIBP Interval
- 13.8.4 Selecting NIBP Start Mode
- 13.8.5 Enabling the NIBP End Tone
- 13.8.6 Setting NIBP Sequence
- 13.8.7 Setting NIBP Measurement Timeout
- 13.8.8 Displaying the NIBP List
- 13.8.9 Correcting the NIBP Measurements
- 13.9 Assisting Venous Puncture
- 13.10 NIBP Maintenance
- 13.11 NIBP Troubleshooting
- 14 Monitoring Invasive Blood Pressure (IBP)
- 14.1 IBP Introduction
- 14.2 IBP Safety Information
- 14.3 Preparing for IBP Monitoring
- 14.4 Measuring ICP Using the Codman ICP Transducer
- 14.5 IBP Display
- 14.6 Changing IBP Settings
- 14.6.1 Changing the IBP Alarm Settings
- 14.6.2 Changing the Pressure Label
- 14.6.3 Setting the Pressure Type for Display
- 14.6.4 Changing the Sensitivity
- 14.6.5 Setting the IBP Waveform
- 14.6.6 Setting the Use PA-D as PAWP Switch (only available for the independent external display)
- 14.6.7 Enabling PPV Measurement
- 14.6.8 Changing the Pressure Unit
- 14.6.9 Overlapping IBP Waveforms
- 14.7 Measuring PAWP (only available for the independent external display)
- 14.8 IBP Troubleshooting
- 15 Monitoring Carbon Dioxide (CO2)
- 15.1 CO2 Introduction
- 15.2 CO2 Safety Information
- 15.3 CO2 Measurement Limitations
- 15.4 CO2 Display
- 15.5 Measuring CO2 Using Sidestream/Microstream CO2 Module
- 15.6 Measuring CO2 Using Mainstream CO2 Module
- 15.7 Changing Settings for All CO2 Modules
- 15.8 Changing Settings for Sidestream and Microstream CO2 Module
- 15.9 Changing O2 Settings (For Sidestream CO2 Module Integrating O2)
- 15.10 Setting the Gas Compensation
- 15.11 Choosing a Time Interval for Peak-Picking
- 15.12 Changing Barometric Pressure
- 15.13 Performing the Leakage Test
- 15.14 CO2 Calibration
- 15.15 Test Method Used to Determine the Respiration Rate Measuring Range
- 15.16 CO2 Troubleshooting
- 15.17 Oridion Information
- 16 Review
- 16.1 Review Overview
- 16.2 Review Page
- 16.2.1 Accessing the Review Page
- 16.2.2 Example Review Page
- 16.2.3 Symbols on Review Pages
- 16.2.4 Common Operations
- 16.2.5 Reviewing the Tabular Trends
- 16.2.6 Reviewing the Graphics Trends
- 16.2.7 Events Review Page
- 16.2.8 Viewing the Full Disclosure
- 16.2.9 OxyCRG Review Page (available for the independent external display)
- 16.2.10 12-Lead ECG Review Page
- 16.3 Minitrends Screen (only available for the independent external display)
- 16.4 Hiding Undesired Review Tabs
- 17 Clinical Assistive Applications (only available for the independent external display)
- 18 Calculation (only available for the independent external display)
- 19 Printing
- 20 Battery
- 21 Care and Cleaning
- 22 Maintenance
- 23 Accessories
- A Product Specifications
- A.1 Monitor Safety Specifications
- A.2 Physical Specifications
- A.3 Environmental Specifications
- A.4 Power Supply Specifications
- A.5 Display Specifications
- A.6 Touchscreen Specifications
- A.7 LEDs
- A.8 Audio Indicator
- A.9 Monitor Interface Specifications
- A.10 Signal Outputs Specifications
- A.11 Data Storage
- A.12 Out-Of-Hospital Transport - Standards Compliance
- A.13 Wi-Fi Specifications
- A.14 Measurement Specifications
- B EMC and Radio Regulatory Compliance
- C Default Settings
- D Alarm Messages
- D.1 Physiological Alarm Messages
- D.1.1 General Physiological Alarm Messages
- D.1.2 Arrhythmia Alarm Messages
- D.1.3 Resp Physiological Alarm Messages
- D.1.4 SpO2 Physiological Alarm Messages
- D.1.5 PR Physiological Alarm Messages
- D.1.6 NIBP Physiological Alarm Messages
- D.1.7 IBP Physiological Alarm Messages
- D.1.8 CO2 Physiological Alarm Messages
- D.2 Technical Alarm Messages
- D.2.1 General Technical Alarm Messages
- D.2.2 ECG Technical Alarm Messages
- D.2.3 Resp Technical Alarm Messages
- D.2.4 SpO2 Technical Alarm Messages
- D.2.5 Temp Technical Alarm Messages
- D.2.6 NIBP Technical Alarm Messages
- D.2.7 IBP Technical Alarm Messages
- D.2.8 CO2 Technical Alarm Messages
- D.2.9 Power Supply Technical Alarm Messages
- D.2.10 Printer Technical Alarm Messages
- D.2.11 Technical Alarm Messages Related to Networked Monitoring
- D.2.12 Other System Technical Alarm Messages
- D.1 Physiological Alarm Messages
- E Electrical Safety Inspection
- F SpO2 Sensor Accuracy
- G Units, Symbols and Abbreviations
B - 4 BeneVision N1 Patient Monitor Operator’s Manual
B.2 Radio Regulatory Compliance
RF parameters
This device complies with Part 15 of the FCC Rules and RSS-210 of Industry Canada. Operation is subject to the
following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
a
The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz
to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0,15 MHz and
80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz
to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz
to 54,0 MHz.
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above,
the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the device.
c
Over the frequency ranges 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Recommended Separation Distances between Portable and Mobile RF, Communications Equipment and This
Equipment
The equipment is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the device as recommended below,
according to the maximum output power of the communication equipment.
Rated Maximum
Output power of
Transmitter Watts
(W)
Separation Distance According to Frequency of Transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.20 2.3
10 3.8 3.80 7.3
100 12 12.00 23
For transmitters at a maximum output power not listed above, the recommended separation distanced in meters (m) can
be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Guidance and Declaration - Electromagnetic Immunity
d
3.5
V
-------
P=
d
3.5
E
-------
P=
d
7
E
---
P=
Radio devices IEEE 802.11b/g/n (2.4G) IEEE 802.11a/n (5G)
Operating Frequency ETSI: 2.4 GHz to 2.483 GHz
FCC: 2.4 GHz to 2.483 GHz
MIC: 2.4 GHz to 2.495GHz
KC: 2.4 GHz to 2.483 GHz
ETSI: 5.15 GHz to 5.35 GHz, 5.47 GHz to 5.725 GHz
FCC:5.15GHz to 5.35GHz, 5.47~5.725GHz to
5.725GHz to 5.825GHz
MIC: 5.15GHz to 5.35GH, 5.47 to 5.725GHz
KC:5.15GHz to 5.35GHz, 5.47 to 5.725GHz, 5.725GHz
to 5.825GHz
Modulation Mode DSSS and OFDM OFDM
Output Power <30dBm (Peak Power)
<20dBm (Average Power)