A - 18 BeneVision N1 Patient Monitor Operator’s Manual
A.14.8 CO
2
Specifications
Accuracy ±2% or ±1 mmHg, whichever is greater (excluding sensor error)
Refreshing rate ≤1s
PPV
Measurement range 0% ~ 50%
Pressure transducer
Excitement voltage 5 VDC, ±2%
Sensitivity 5 V/V/mmHg
Zero adjustment range ±200 mmHg
Impedance range 300 to 3000
Volume displacement <0.04 mm
3
/100 mmHg
Alarm limit
Range (mmHg)
Step (mmHg)
Sys High IBP ≤ 50: (low limit + 2) to 50
IBP > 50: (low limit + 5) to 355
IBP ≤ 50: 1
IBP > 50: 5
Mean High
Dia High
Sys Low IBP ≤ 50: - 49 to (high limit - 2)
IBP > 50: 50 to (high limit - 5)
Mean Low
Dia Low
Art-S Extreme High (High limit+5) to 360 IBP ≤ 50: 1
IBP > 50: 5
Art-M Extreme High
Art-D Extreme High
Art-S Extreme Low -50 to (low limit-5)
Art-M Extreme Low
Art-D Extreme Low
Measurement mode Sidestream, microstream, mainstream
Technique Infrared absorption
Zero RR delay 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit Range Step
EtCO
2
High (low limit + 2) to 99 mmHg 1 mmHg
EtCO
2
Low 1 to (high limit - 2)mmHg
FiCO
2
High 1 to 99 mmHg
EtO
2
High (low limit + 2%) to 100% 1%
EtO
2
Low 18% to (high limit - 2)%
FiO
2
High (low limit + 2%) to 100%
FiO
2
Low 18% to (high limit - 2)%