Operator's Manual

Table Of Contents
BeneVision N1 Patient Monitor Operator’s Manual A - 15
PR from Nellcor SpO
2
Module
PR from Masimo SpO
2
Module
PR from NIBP Module
PR from IBP Module
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time ≤30 s (normal perfusion, no disturbance, PR value sudden change from 25 to
250 bpm)
Accuracy 20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate ≤1 s
Measurement range 25 to 240 bpm
Resolution 1 bpm
Response time ≤20 s (normal perfusion, no disturbance, PR value sudden change from 25 to
220 bpm)
Accuracy ±3 bpm (without motion)
±5 bpm (with motion)
Refreshing rate ≤2 s
Measurement range 30 to 300 bpm
Resolution 1 bpm
Accuracy ±3 bpm or ±3%, whichever is greater
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater