Operator's Manual

Table Of Contents
22 - 2 BeneVision N1 Patient Monitor Operator’s Manual
22.3 Maintenance and Testing Schedule
Follow the maintenance and testing schedule or local regulations to perform testing and maintenance. Make
sure to clean and disinfect the equipment before taking any tests and maintenance
The following table lists the maintenance and testing schedule:
22.4 Checking Version Information
You may be asked for information on monitor and module version.
To view system software version information, select the
Main Menu quick key from the System column select
Version.
You can also view more version information by following this procedure
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Version tab.
You can check system software version, module hardware and software version, and firmware version.
22.5 Testing Methods and Procedures
Except the following maintenance tasks, all other test and maintenance tasks should be performed by Mindray-
qualified service personnel only.
Regular check, including visual inspection and power-on test
Printer tests
Battery check
If your monitor needs a safety test and performance test, contact the service personnel.
Test/Maintenance Item Recommended Frequency
Performance Tests
Visual inspection Every day, before first use.
Measurement module performance test and
calibration
1. If you suspect that the measurement values are incorrect.
2. Follow any repairs or replacement of relevant module.
3. Once a year for CO
2
tests.
4. Once every two years for other parameter module performance
tests.
Analog output test If you suspect that the analog output function does not work
properly.
Defibrillation synchronization test If you suspect that the defibrillation synchronization function does
not work properly.
Electrical Safety Tests
Electrical safety tests Once every two years.
Other Tests
Power-on test Before use.
Network printer tests 1. When first installed.
2.Followany repair or replacement of the printer.
Battery check Functionality test 1. When first installed.
2. When battery is replaced.
Performance test Every three months or if the battery runtime reduced significantly.