BeneVision N1 Patient Monitor Operator’s Manual 19 - 3
Select the Main Menu quick key from the Parameters column select Setup select the desired
parameter
select the Alarm tab.
2. Switch on
On/Off and Alarm outputs for desired parameters.
19.7 Stopping a Printing Task
To stop a printing task, follow this procedure:
1. Select the
Main Menu quick key from the Report column select Print Queue.
2. Select desired printing tasks and then select
Delete. Selecting Delete All to stop all the printing tasks.
19.8 Setting Reports
This section focuses on how to set ECG reports, realtime reports, tabular trends reports, and graphic trends
reports.
19.8.1 Setting ECG Reports
To set ECG reports, follow this procedure:
1. Select the
Main Menu quick key from the Report column select Normal Report.
2. Select ECG Report.
3. Set
the desired options. The following table only lists some of the options.
NOTE
When ECG Lead Set is set to 3-Lead, ECG report cannot be printed.
Menu item Function Description
Auto Interval Defines the spacing
between the ECG
waveforms
on a
printout
On: automatically adjusts the space between waveforms to
avoid overlapping.
Off: each waveform area has the same sizeon a printout.
Note: This setting is only relevant when
12×1 is selected for 12-Lead Format.
12-Lead Format Select the format of 12-
lead ECG waveforms
on a printout.
12×1: displays 12-lead ECG waveforms on one page in one
column.
6×2: displays 12-lead ECG waveforms on one page in two
columns, with 6 lines in each column.
6×2+1: displays 12-lead ECG waveforms on one page in two
columns, with 6 lines in each column, and one rhythm lead
waveform at the bottom.
3×4+1: displays 12-lead ECG waveforms on one page in 4
columns, with 3 lines in each column, and one rhythm lead
waveform at the bottom.
3×4+3: displays 12-lead ECG waveforms on one page in 4
columns, with 3 lines in each column, and three rhythm lead
waveforms at the bottom.
Rhythm Lead 1
Rhythm Lead 2
Rhythm Lead 3
Select the lead that will
be used as Rhythm
Lead 1, 2, or 3.
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Note: This setting is only relevant when
6×2+1,3×4+1, or 3×4+3 is selected for 12-Lead
Format
.
Format sequence Select the recording
method of ECG report
generated by auto
measurement
Sequential: 12-lead ECG data are recorded sequentially and
displayed in 3 lines and 4 columns with 2.5 seconds of ECG data
for each column.
Simultaneous: Record simultaneous 12-lead ECG data.