Operator's Manual

Table Of Contents
19 - 2 BeneVision N1 Patient Monitor Operator’s Manual
19.3.2 Setting Patient Information on Other Reports
To set the patient information you want to display on other reports, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Print tab.
3. Select the
Report Layout tab.
4. Select the desired items under
Report Name. N/A indicates that this item is not displayed on a report.
19.4 Setting End Case Reports
All normal reports can be set as end case reports. If you enable Print End Case Report when you discharge a
patient, the monitor will automatically print the end case reports.
To set a normal report as an end case report, follow this procedure:
1. Select the
Main Menu quick key from the Report column select End Case Report.
2. Select the desired report tab, for example
ECG Report.
3. Switch on
Set As End Case Report to set the selected report to the end case report.
19.5 Manually Starting a Printing Task
You can start a printing task manually.
19.5.1 Starting Printing from the Current Page
From the current page, select the button, if available, to start printing.
19.5.2 Printing Realtime Reports
Select the quick key to print a realtime report. You can also print a realtime report from the Normal
Report page. For more information, see 19.5.3 Printing Normal Reports.
19.5.3 Printing Normal Reports
Normal reports refers to the following types of reports:
ECG Report
Realtime Report
Tabular Trends Report
Graphic Trend Report.
To print normal reports, follow this procedure:
1. Select the
Main Menu quick key from the Report column select Normal Report.
2. Select the desired report tab.
3. Check the settings.
4. Select
Print.
19.6 Automatically Printing Reports
When a parameter alarm switch is set to on and an alarm is triggered for this parameter, you can set a printer to
start alarm printing automatically.
To do so, follow this procedure:
1. Access alarm related tabs such as the
Alarm tab for a parameter in one of the following ways:
Select the parameter or waveform area of the desired parameter select the Alarm tab.