Operator's Manual

Table Of Contents
BeneVision N1 Patient Monitor Operator’s Manual 16 - 5
Earlier events will be overwritten by later ones if the capacity is reached.
16.2.7.1 Entering the Events Review Page
To enter the events review page, select the Main Menu quick key from the Review column select Events.
The
Event page displays event list. Events are displayed in descending chronological order. The most recent
event is displayed at the top.The number of asterisk symbols before an event indicate alarm priorities. Different
color blocks are displayed on the left of each event to indicate different event types.
Red: high priority alarm event
Yellow: medium priority alarm event
Cyan: low priority alarm event
Green: manual event
White: operation-related event
16.2.7.2 Configuring the Filter
You can filter events by time, alarm priority, alarm category, or parameter group. To configure the filter, follow
this procedure:
1. Enter the
Events page.
2. Select
Event List.
3. Select
Filter Setup and set the desired filter criterion.
16.2.7.3 Editing Events
To edit events, follow this procedure:
1. Enter the
Events page and tick off the desired events.
2. Select
to edit the selected events.
Lock: manually lock the event. Locked events cannot be deleted.
Note: enter comments for the event.
16.2.7.4 Viewing Event Details
To view waveforms and parameter values at the event time, follow this procedure:
1. Enter the
Events page.
2. Select
Detail.
To display beat labels on the first ECG waveform, switch on
Beat Anno. The white beat labels indicate heart
beats classification and may explain suspected, missed, or false arrhythmia calls. Heart beats are classified as
follows:
N = Normal
V = Ventricular ectopic
S = Supraventricular premature
P = Paced
L = Learning
? = Insufficient information to classify beats
I = Inoperative (for example, Lead Off)
M = Missed beat
16.2.7.5 Printing Event Reports
To print event reports, follow this procedure:
1. Enter the events review page.