1. Select the Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab Other tab.
3. Set
CO2 Flow Rate For Neo.
15.9 Changing O
2
Settings (For Sidestream CO
2
Module Integrating O
2
)
15.9.1 Changing O
2
Alarm Settings
To change the O
2
alarm settings, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Alarm tab.
3. Set alarm properties as desired.
15.9.2 Changing the O
2
Unit
To change the O
2
unit, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Unit tab.
3. Set
O2 Unit.
15.9.3 Setting the O
2
Waveform
To set the O
2
waveform, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set
Speed and O2 Scale of the O
2
waveform.
15.10 Setting the Gas Compensation
The presence of interfering gas affects the CO
2
measurement. To get the best possible measuring result, it is
needed to set the gas compensation. The configured concentration of the interfering gas should be in
accordance with its actual proportion.
For the microstream CO
2
module, gas compensations are not required.
WARNING
Make sure to use the appropriate compensations. Inappropriate compensations may cause
inaccurate measurement values and result in misdiagnosis.
For the sidestream CO
2
module, follow this procedure to set the gas compensation:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set the compensation according to the actual condition.
For the mainstream CO
2
module, follow this procedure to set the gas compensation:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set the following compensation according to the actual condition.
Balance Gas