15 - 2 BeneVision N1 Patient Monitor Operator’s Manual
15.2 CO
2
Safety Information
WARNING
Route all tubing away from the patient’s throat to avoid strangulation.
CAUTION
Remove the airway sample line from the patient’s airway while nebulized medications are being
delivered.
EtCO
2
values measured from the CO
2
module may differ from those of from the blood gas analysis.
Avoid mechanical shock to the sidestream CO
2
module
NOTE
The CO
2
module automatic suppress physiological alarms until breathing waves have been
detected. Make sure that a patient is properly connected when monitoring with the CO
2
module.
15.3 CO
2
Measurement Limitations
The following factors may influence the measurement accuracy:
Leaks or internal venting of sampled gas
Mechanical shock
Cyclic pressure up to 10 kPa (100 cmH
2
O)
Other sources of interference, if any
Measurement accuracy of the sidestream CO
2
module may be affected by the breath rate and inspiration/
expiration (I/E) ratio. Measurement accuracy of the microstream CO2 module may be affected by the breath
rate.For more information, refer to
A.14.8CO
2
Specifications.