BeneVision N1 Patient Monitor Operator’s Manual 13 - 5
2. Select Initial Pressure, and then select the appropriate setting.
13.8.3 Setting the NIBP Interval
For auto NIBP measurement, you need to set the interval between two NIBP measurements. To set the NIBP
interval, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Set
Interval. Selecting Manual switches to manual mode.
13.8.4 Selecting NIBP Start Mode
Start mode defines how NIBP auto mode works. To set the start mode, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Set
Start Mode.
Clock: after the first measurement, the monitor automatically synchronizes NIBP automatic
measurements with the real time clock. For example, if
Interval is set to 20 min, and you start NIBP
auto measurement at 14: 03, the next measurement will be taken at 14: 20, and then at 14:40, 15:00,
and so on.
Interval: after the first measurement, the monitor automatically repeats measurements at set
interval. For example, if
Interval is set to 20 min, and you start NIBP auto measurement at 14:03, the
next measurement will be taken at 14:23, and then at 14:43, 15:03, and so on.
13.8.5 Enabling the NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP End Tone is off by
default. To switch on the NIBP end tone, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Switch on
NIBP End Tone.
13.8.6 Setting NIBP Sequence
NIBP sequence measurement can have up to five phases: A, B, C, D, and E. You can individually set the duration
and interval of each phase.
To set NIBP sequence, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Select the
Sequence tab.
3. Set
Duration and Interval of each phase.
13.8.7 Setting NIBP Measurement Timeout
NIBP measurements become outline fonts after a preset time. This avoids older NIBP values being
misinterpreted as current measurements. To set the timeout period, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab select Other.
3. Set
NIBP Timeout.
13.8.8 Displaying the NIBP List
To display multiple sets of the most recent NIBP measurements, follow this procedure:
1. Access
Tile Layout by either of the following ways:
Select the Screen Setup quick key select the Tile Layout tab.