13 - 4 BeneVision N1 Patient Monitor Operator’s Manual
4. Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass
unrestricted through the tubing.
CAUTION
A wrong cuff size and a folded or twisted bladder can cause inaccurate measurements.
Do not touch or apply external pressure against the cuff and air tubing during NIBP measurement.
This may cause inaccurate blood pressure values.
Use care when placing the cuff on an extremity used for monitoring other patient parameters.
13.7 Starting and Stopping NIBP Measurements
Start and stop NIBP measurement by selecting the NIBP quick keys or from the NIBP menu.
13.8 Changing NIBP Settings
13.8.1 Setting the NIBP Alarm Properties
To set the NIBP alarm properties, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Select the
Alarm tab.
3. Enter the password if required. For more information, refer to
6.9.3 Selecting Password for User
Authentication
.
4. Set alarm properties as desired.
13.8.2 Setting the Initial Cuff Inflation Pressure
To set initial cuff inflation pressure, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
Task By Quick Key From NIBP menu
Start a manual measurement NIBP Start/Stop quick key Start NIBP button
Start auto NIBP series
NIBP Start/Stop quick key
Make sure to set the
Interval before
starting the auto NIBP.
Setup tab set Interval Start
NIBP
button
NIBP Measure quick key select
interval
Start
NIBP sequence measurement
NIBP Measure quick key
Sequence
Sequence
tab set NIBP sequence
Start NIBP button
Start STAT measurement
NIBP Measure quick key
STAT
STAT
button
Stop the current NIBP
measurements
NIBP Start/Stop quick key Stop NIBP button
End auto NIBP series or NIBP
Sequence
/Stop All button
Stop STAT measurement and end
series
NIBP Start/Stop quick key Stop NIBP or Stop All button