Operator's Manual

Table Of Contents
8 - 18 BeneVision N1 Patient Monitor Operator’s Manual
3. From the bottom of the menu, select ST Graphic.The following figure shows the ST Graphic when ST
Alarm Mode is set to Absolute. The height of the bar indicates the ST value of corresponding ST lead. The
color of the bar indicates ST alarm status: green indicates that corresponding ST value is within alarm limits;
cyan, yellow and red indicate that the ST value exceeds the alarm limits. The color matches ST alarm
priority.
The following figure shows the ST Graphic when ST Alarm Mode is set to Relative. The height of grey bar
indicates the baseline ST value and the green bar (cyan, yellow or red if an alarm occurs) indicates ST.
8.8 Changing ST Settings
8.8.1 Setting ST Alarm Properties
To set ST alarm properties, follow this procedure:
1. Select the ST numeric area, ECG numeric area or ECG waveform area to enter the ECG menu.
2. Select the ST tab Alarm tab.
3. Set ST Alarm Mode to Absolute or Relative.
Absolute: you can separately set the alarm properties for each ST alarm.
Relative: you can set the alarm properties for ST Single and ST Dual alarms.
4. Set ST alarm properties.
8.8.1.1 Setting the ST Alarm Delay Time
To set the ST alarm delay time, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab select the Other tab.