Operator's Manual

Table Of Contents
8 - 16 BeneVision N1 Patient Monitor Operator’s Manual
2. Select the ST tab select the Setup tab.
3. Switch on ST Analysis.
Reliable ST monitoring cannot be ensured under the following situations:
You are unable to get a lead that is not noisy.
Arrhythmias, such as atrial fib or flutter, cause irregular baseline.
The patient is continuously ventricularly paced.
The patient has left bundle branch block.
In these cases, you may consider switching off ST monitoring.
8.7.3 Displaying ST Numerics
To display ST numerics and Segments, follow this procedure:
1. Access Tile Layout by either of the following ways:
Select the Screen Setup quick key select the Tile Layout tab.
Select Main Menu quick key from the Display column select Tile Layout.
2. Click the numeric area where you want to display the ST numerics, and then select ECG ST.
The display of ST parameters area is different according to the lead type:
When you are using the 3-lead ECG leadwires, the ST numeric area does not display. A ST value displays in
the ECG numeric area.
When you are using the 5-lead ECG leadwires, the ST numeric area displays 7 ST values: ST-I, ST-II, ST-III, ST-
aVR, ST-aVL, ST-aVF, ST-V.
When you are using the 6-lead ECG leadwires, the ST numeric area displays 8 ST values: ST-I, ST-II, ST-III, ST-
aVR, ST-aVL, ST-aVF, ST-Va, ST-Vb.
When you are using the 12-lead ECG leadwires, the ST numeric area displays12 ST values: ST-I, ST-II, ST-III,
ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-V2, ST-V3, ST-V4, ST-V5, ST-V6.
This example shows the ST numeric area when 5-lead ECG cable is used. Your monitor screen may look slightly
different:
8.7.4 Displaying ST Segments in the Waveform Area
You can display ST segments in the waveform area. To do so, follow this procedure:
1. Access Tile Layout by either of the following ways:
Select the Screen Setup quick key select the Tile Layout tab.
Select Main Menu quick key from the Display column select Tile Layout.
(1) Parameter label (2) ST unit
(3) ST alarm off symbol (4) Lead labels
(5) ST numerics: a positive value indicates ST segment elevation, and a negative value indicates ST
segment depression.
(4) (5)
(1)
(2)
(3
)