8.6.4.5 Setting Thresholds for PVC-Related Alarms
PVC-related alarms are detected on the basis of the current PVC rate and the number of consecutive PVCs.
To set the required thresholds for PVC-related alarms, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG menu.
2. Select the Arrhythmia tab select the More Threshold tab.
3. Enter the password if required. For more information, refer to 6.9.3 Selecting Password for User
Authentication.
4. Adjust V-Tach PVCs, V-Tach Rate, V Brady PVCs, and V Brady Rate to set the threshold of desired PVC-
related alarms.
The following figure illustrates the conditions under which PVC alarms will be generated if V-Tach PVCs is set to
6, V-Tach Rate is set to 130, V Brady PVCs is set to 5, and V Brady Rate is set to 40.
If both V-Tach PVCs and V-Tach Rate are greater than or equal to the limits, a V-Tach alarm is generated.
Extreme Brady 15bpm to 115 bpm
Multif PVCs Window 3 beats to 31 beats
V-Tach Rate 100 bpm to 200 bpm
V Brady Rate 15 bpm to 60 bpm
V-Tach PVCs 3 beats to 99 beats
V Brady PVCs 3 beats to 99 beats
PVCs/min 1 to 100
Pauses/min 1 to 15
Pause Threshold 1.5sec, 2.0sec, 2.5sec, 3.0sec
AF/Irr Rhy End Time 0, 1 min, 2 min, 3 min, 4 min, 5 min, 10 min, 15 min, 30 min