User's Manual

Viewing Alarm List Alarms
6 - 6 TM80/TM70 Telemetry Monitor Operator’s Manual
6.5.5 Alarm Status Symbols
The TM80/TM70 still uses the following symbols indicating the alarm status:
■ : indicates that all the alarms are paused.
■ : indicates the technical alarm audio is turned off.
■ : indicates the alarm system is reset.
6.6 Viewing Alarm List
1. Tap the message area on the main screen. The [Alarm List] menu is displayed.
2. In the [Physiological Alarms] section, you can view the number of physiological
alarms and alarm messages.
3. In the [Technical Alarms] section, you can view the number of technial alarms and
alarm messages.
6.7 Configuring the Alarms
You can set alarm-related items such as alarm properties, alarm volume, and alarm setup
menu. For how to set alarm volume and alarm setup menus, refer to "Configuring the
Audio Volume" on page 4 - 4 and "Configuring the Alarms Menu" on page 13 - 4
respectively.
6.7.1 Changing Alarm Properites
You can change alarm properties for parameters collectively or individually.
To change alarm properties, follow this procedure:
1. Enter the [Alarm Limit] menu in either of the following ways:
◆ Press the key to enter the main menu →tap [Alarms] →tap [Alarm
Limits].
◆ Tap the desired digital area or waveform area to enter corresponding
parameter setup menu → from the [Alarms] section tap alarm-related
option.
2. In the [Alarm Limits] section, switch on or off a parameter alarm.
3. Tap the desired parameter to enter its alarm setup menu.
4. In the [Limits] section, tap the alarm limit which you want to change.
5. Enter the new alarm limt.
6. Tap to return to the parameter alarm setup menu.
7. In the [Responses] section, set the desired alarm priority.