User's Manual

Safety Information Safety
1 - 4 TM80/TM70 Telemetry Monitor Operator's Manual
1.1.2 Cautions
WARNING
• Only use parts and accessories specified in this manual.
• Route, wrap and secure the cables to avoid inadvertent disconnection,
stumbling and entanglement
CAUTION
• Do not let the display of the TM80/TM70 directly touch the patient when
the display is on.
• When the Central Monitoring System presents the alarm “Offline” for
TM80/TM70, the setting being performed on the TM80/TM70 may not
be transferred to the Central Monitoring System. Check the patient con-
dition and the settings on the Central Monitoring System.
• When disposing of the packaging material, be sure to observe the appli-
cable waste control regulations and keep it out of children’s reach.
• The patient should be required to move in a specified area. If the patient
is at the edge of or outside the network coverage range, unstable net-
work connection may compromise the monitoring performance. The
patient’s location is of vital importance for the TM80/TM70. If a life-
threatening situation occurs for a patient, this patient must be located
and found by medical staff immediately.
• Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. For this reason make sure that all exter-
nal equipment operated in the vicinity of the equipment comply with
the relevant EMC requirements. Mobile phone, X-ray equipment or MRI
equipment are a possible source of interference as they may emit higher
levels of electromagnetic radiation.
• Always install or carry the TM80/TM70 properly to avoid damage caused
by drop, impact, strong vibration or other mechanical force.
• Dry the TM80/TM70 immediately in case of rain or water spray.