Service manual

Trio™ Service Manual 0070-10-0591-01 2 - 3
Calibration/Maintenance Calibration and System Checks
FIGURE 2-2 Maintenance Menu (below S/N MC15000-XX)
2.4.2.1 Calibrations
Calibrations should be performed at least once a year. Calibrations should also be
performed after any preventive maintenance or repair of the Trio Monitor.
From the MAINTENANCE menu make the following selections to perform each specific
calibration or check.