S e r v i ce M a n u a l 0070-01-0591-02_revB_Trio srv color.
S e r v i ce M a n u a l
Trio™ is a U.S. trademark of Mindray DS USA, Inc. Velcro® is a registered trademark of Velcro Industries B.V. Navigator™ is a U.S. trademark of Mindray DS USA, Inc. Masimo SET® is a U.S. registered trademark of Masimo Corp. Copyright © Mindray DS USA, Inc., 2008. All rights reserved. Contents of this publication may not be reproduced in any form without permission of Mindray DS USA, Inc.
Table of Contents Foreword ....................................................................................................................................................... iii Warnings, Precautions And Notes .................................................................................................................... iii Warning........................................................................................................................................................
Table of Contents Removal of Power Supply Assembly ........................................................................................................... 4 - 15 Removal of PCB Chassis Rear Panel Plate ................................................................................................... 4 - 15 Removal of NIBP/IBP PCB Mounting Plate .................................................................................................. 4 - 16 Removal of Handle ...................................
Foreword Introduction Foreword This service manual gives a detailed description of the Trio Portable Patient Monitor, including, circuit descriptions, test procedures and a spare part listing. This manual is intended as a guide for technically qualified personnel during repair, testing or calibration procedures. Warnings, Precautions And Notes Please read and adhere to all warnings, precautions and notes listed here and in the appropriate areas throughout this manual.
Introduction Warning This page intentionally left blank.
1.0 Theory of Operation 1.1 Introduction The Trio portable patient monitor uses a parameter module as the basis for acquiring patient data. The results are transmitted to the main control board to process and display the data and waveforms. CPU board commands and status messages of modules are transmitted via databus. The structure of the entire system is shown in the figure below.
Introduction Theory of Operation As shown in the above figure, the four parameter modules execute real-time monitoring of NIBP, SpO2, ECG/RESP/TEMP and IBP through the use of blood pressure cuffs and patient cables. The patient data is transmitted to the CPU board for display. When required, data may be printed out via the recorder.
Theory of Operation 1.2 Hardware Overview Hardware Overview TFT Display 834 inchs 800 X 600 FAN J2 Alarm LED P5 P1 J9 A4(TFT_DIGTAL) J6 P10 P2 (CRT) P3 (FOR 9000 P12 VGA) X16 Power Supply PCB J3 Serial interface X15 X14 Main Power Input Serial Coverter J1 P.C.B J5 J4 P11 J2 J7 Speaker Host P .C.B. X 1 Recorder Module X 2 P17 (FOR 509C) TO X4 J8 P7(BDM ) P13 Battery Key & Alarm P .C.B. P15 P16 P6 P9 P14 P8 From J2 X4 X 3 Recorder P.S.
Hardware Overview Theory of Operation 5VDC-DC converter AC input Voltage test AC/DC REC POWER SOURCE Battery and Charging Management Circuit Power on/off control circuit 12V output FIGURE 1-3 Block diagram of Trio power supply board Key Test Points 1-4 NO. NAME LOCATION FUNCTION 1 Rectified voltage C12 Primary rectified voltage, range: 107~354 V 2 RTN1 C12 negative electrode Primary ground 3 Driving waveform Q1.1 There is a driving waveform of about 100 KHZ between Q1.
Theory of Operation 1.2.2 Hardware Overview Power Supply Board (Lithium Ion Battery) P/N 0671 -00-0051 Trio Power Supply board specifications: • AC input voltage:110~240VAC+10% • AC input current: <1.6A • AC voltage frequency: 50/60+3HZ • Two-way output voltage: 5V/12V, normal working current is 1.3A for 5V, 1.3A for 12V. • Two-way output voltage has functions of short-circuit, over-current and over-voltage protection. • The power board has reset function.
Hardware Overview Theory of Operation Key Test Points 1.2.3 NO. NAME LOCATION FUNCTION 7 12V C41 positive electrode 12V output, voltage range is 11.0~13.0V 8 5V C58 positive electrode 5V output, voltage range is 4.75~5.25V 9 Feedback voltage R37 positive electrode There is a DC waveform of about 2.5V between R37 CPU Board (Main Control Board) P/N 0671-00-0056 or P/N 0671-00-0236 1.2.3.
Theory of Operation Hardware Overview 2 RTC/E PROM DRAM Flash/SRAM Display driving circuit CPU Network controller Interrupt management circuit I/O serial interface Audio alarm/spare battery FPGA FIGURE 1-5 Block Diagram of Trio CPU board 1.2.3.2 Detailed Description 3.3 V low-voltage power supply component is used. The external power is 5 V, which is converted by the DC/DC converter into 3.3 V and 2.5 V, the latter voltage being especially used for FPGA.
Hardware Overview Theory of Operation Display The resolution is 800 x 600. Frequency is 38 MHz. It works in an appropriate SVGA mode. VRAM adopts 16-bit structure and is divided into an alphanumeric character screen and a waveform screen. To the left of the alphanumeric character screen is the corresponding waveform screen. The character screen is used to display data and flashing alarming parameters. The user can select the color of the waveform and alphanumeric characters for each parameter.
Theory of Operation 1.2.4 Hardware Overview Keypad Board P/N 0671-00-0237 CPU (AT89C2051) Watchdog RAM 128 x 8 BUTTON button and encoder scan circuit button signal input FLASH 4KX8 serial communication Main control board Alarm indicator control circuit ENCODER Sound Effect Control Volume Control Lowpass and Bandpass Filter Power Amplifier LM386 speaker FIGURE 1-6 Keypad Block Diagram 1.2.4.1 Detailed Description This circuit has three main parts: 1.
Hardware Overview 1.2.5 Theory of Operation Keypad Board P/N 0671-00-0058 LED KEY CPU (AT89C51) SPEAKER ENCODER Audio Process Circuit RAM 128 X 8 FLASH 4K X 8 Watchdog Main Control board (Host CPU Board) FIGURE 1-7 Keypad block diagram 1.2.5.1 Detailed Description This module detects key and encoder input signals, converts them into code then sends these coded signals to the main board.
Theory of Operation Hardware Overview Key Test Points 1.2.6 NO. NAME LOCATION FUNCTION 1 VCC P4.4 Power input, range: 4.8~5.2V 2 GND P4.5 Power supply and signal ground 3 RST U1.10 CPU reset signal. At low level(<0.3V) when operating normally 4 Crystal oscillator X1.1,X.2 CPU crystal oscillator. Sine wave signal (1.5~3.
Hardware Overview Theory of Operation Audio Process Circuit Generates audio signals to drive the speaker. Watchdog • Upon power-up, supply Reset signal to CPU • Provide functions of Watchdog Timer Output and voltage detection. Key test points 1.2.7 NO. NAME LOCATION FUNCTION 1. 5V/5B J5 pin 1 Power input, range: 4.0~5.5V 2. GND J5 pin 2 Power supply and signal ground 3. RST J5 pin 3 CPU reset signal. At low level(<0.8V (during normal operation) 4.
Theory of Operation Hardware Overview CPU System The CPU system is the core of the drive board. Its task is to receive the data from the host and generate lattice messages after calculation using a specified algorithm. These messages are then sent to the thermal head for printing. The CPU system can simultaneously collect data from both the thermal head and the drive board and display data sent to the host. Power Conversion The recorder requires the system to provide two voltages: 12 V and 5 V.
Serial Interface Converter Board 1.3 Theory of Operation Serial Interface Converter Board The Serial Interface Converter Board is used to convert the TTL level (5V) to RS232 level. P5 P1 J2 Serial Coverter P.C.B J1 Serial interface P4(TFT_DIGTAL ) P10 P2(CRT) P3(FOR 9000 VGA) P12 P11 P7(BDM ) Host P .C .B.
Theory of Operation Parameter Circuit Descriptions 1.4 Parameter Circuit Descriptions 1.4.1 ECG The main functions concerning ECG are: • Lead: 3-lead, 5-lead • Lead Method; I, II, III, avR, avL, avF, V, CAL • Floating Input • Right-Leg Drive • Lead-off Detection The ECG circuit is responsible for processing the ECG signals of human body. The circuit consists of following parts: Input Circuit: The ECG electrodes are connected into the circuit through the cable.
Parameter Circuit Descriptions 1.4.3 Theory of Operation NIBP The monitor measures non-invasive blood pressure using the oscillometric method. Detailed measurement procedures follows: 1. Inflate the cuff encircled around the upper arm until the pressure in the cuff blocks the blood flow in the artery of the upper arm. 2. Then deflate the cuff according to the requirement of the algorithm. 3.
Theory of Operation Parameter Circuit Descriptions The sensor measurement wavelengths are nominally 660 nm for the Red LED and 940 nm for the Infrared LED. Maximum optical power output for LED is 4 mW. The amount of light transmitted depends on many factors, most of which are constant. However, one of these factors, arterial blood flow, varies with time because it is pulsating. By measuring the light absorption during a pulsation, it is possible to derive the oxygen saturation of the arterial blood.
Parameter Circuit Descriptions Theory of Operation This page intentionally left blank.
2.0 Calibration/Maintenance 2.1 Calibration Introduction The following procedures are provided to verify the proper operation of the Trio Monitor. A menu driven interface is used to execute all verification tests. 2.2 Warnings and Guidelines In the event that the instrument cover is removed, observe these following warnings and general guidelines: 1. Do not short component leads together. 2. Perform all steps in the exact order given. 3. Use extreme care when reaching inside the opened instrument.
Calibration and System Checks Calibration/Maintenance 2.4 Calibration and System Checks 2.4.1 Device Appearance and Installation Checks Inspect the Trio Monitor to ensure that: • The outer housing of the device is clean and has no scratches or cracks • When shaking the device, there are no loose components • All keys are smooth and free for operation • Labels are complete, sufficient and accurate • Standard configuration is complete and all connectors are installed securely 2.4.
Calibration/Maintenance Calibration and System Checks FIGURE 2-2 Maintenance Menu (below S/N MC15000-XX) 2.4.2.1 Calibrations Calibrations should be performed at least once a year. Calibrations should also be performed after any preventive maintenance or repair of the Trio Monitor. From the MAINTENANCE menu make the following selections to perform each specific calibration or check.
Calibration and System Checks Calibration/Maintenance IBP Pressure Cal The purpose of this calibration is to ensure that the system gives you accurate measurements. FIGURE 2-3 IBP Calibration WARNING: Never perform this procedure while patient is being monitored. 1. Connect a Pressure Transducer Interface Cable from a Pressure Transducer (Figure 2-3) to the IBP socket on the right side panel of the Trio 2. Open 3-Way Stopcock to atmosphere. 3.
Calibration/Maintenance Calibration and System Checks FIGURE 2-4 IBP Pressure Calibration Menu 10. Select CALIBRATE. The monitor will perform the calibration and display one of the following calibration completion messages. Trio™ Service Manual MESSAGE DESCRIPTION SUCCESSFUL CALIBRATE Indicates Blood Pressure was calibrated successfully. No further action required. SENSOR OFF, FAIL Check that sensor is connected, then proceed with calibration. If message does not clear contact Customer Service.
Calibration and System Checks Calibration/Maintenance NIBP CAL Connect a test chamber (P/N 0138-00-0001-03) and a calibrated digital mercury manometer with bulb via T fitting to the NIBP quick connect fitting of the Trio Monitor (Figure 2-5.) 1. From the MAINTENANCE menu select NIBP CAL. The selected tile will now read STOP CAL. The user may select this tile at any time to stop the test. 2. Using the bulb, inflate the pneumatic system so that the digital mercury manometer reads 0, 50 and 200 mmHg in turn.
Calibration/Maintenance Calibration and System Checks NIBP Pneumatic WARNING: The NIBP pneumatic test (specified in the EN 1060-1 standard) is used to determine if there are air leaks in the NIBP airway. If the system displays the prompt that the NIBP airway has air leaks, please contact the manufacturer for repair. 1. Connect a test chamber (P/N 0138-00-0001-03) to the NIBP quick connect fitting on the Trio Monitor (Figure 2-7). 2. From the MAINTENANCE menu, select NIBP PNEUMATIC.
Calibration and System Checks Calibration/Maintenance ECG Cal 1. Set ECG lead to I and ECG size to x1. 2. From The MAINTENANCE menu select ECG CAL. The message CALIBRATION MODE will appear in red at the lower left of the ECG waveform area. The highlighted tile will now read STOP CAL. The user may select this tile at any time to end the test. 3. A 1mV peak to peak square wave will appear in the ECG waveform area. This is the Cal Pulse. 4. Measure the amplitude of the Cal Pulse with a metric ruler.
Calibration/Maintenance Calibration and System Checks Error Log: This menu will display any system errors which may have occurred during normal operation of unit. SCROLL or PRINT this information by selecting the respective tiles FIGURE 2-9 Error Log Menu* Factory: This menu allows the unit to be preconfigured at the factory.
Calibration and System Checks Calibration/Maintenance FIGURE 2-10 Factory Menu Demo: This menu allows the unit to operate in Demonstration Mode. When the unit is operating in Demonstration mode the word DEMO will be displayed on the screen at all times. This menu is enabled with an access code and should be used for demonstration purposes only.
Calibration/Maintenance 2.4.3 Calibration and System Checks Safety Tests Equipment: Safety Analyzer (Dempsey Model 431 or equivalent). 2.4.3.1 Leakage Current Tests Case Leakage 1. Plug the line cord of the unit into the safety analyzer. 2. Connect the case ground lead of the analyzer to the equipotential lug of the monitor (at rear of monitor). 3. Perform the tests under the following conditions: a. Case Grounded: Normal polarity Polarity with open neutral b.
Testing Each Parameter 2.5 Calibration/Maintenance Testing Each Parameter Testing each parameter annually ensures the accuracy of the Trio Monitor. The following tests should be used for verification of operation only. The following tests should not be considered calibration procedures. 2.5.1 ECG and RESP 2.5.1.1 Test Equipment • Patient Simulator 2.5.1.2 Test Procedures 1. Use a 5L patient cable and lead wires to connect a calibrated patient simulator to the ECG connector of Trio. 2.
Calibration/Maintenance 2.5.2.2 Testing Each Parameter Test Procedures 1. Use polyurethane tubing to connect the Trio Monitor to a calibrated NIBP simulator and an NIBP Dummy Cuff/Test Chamber (P/N 0138-00-0001-03) via a T fitting. 2. Select Adult mode for both the NIBP simulator and the Trio Monitor. 3. Select a group of blood pressure values within the measurement range on the NIBP simulator, such as: Systolic = 90 Mean = 70 Diastolic = 60 4. From the NIBP SETUP menu set INTERVAL to 1 min. 5.
Testing Each Parameter Calibration/Maintenance 2.5.4 TEMP 2.5.4.1 Test Equipment • Patient Simulator 2.5.4.2 Test Procedures 1. Connect a 400 series TEMP test cable to the TEMP connector of the Trio Monitor. 2. Connect the 400 series test cable to a calibrated Patient Simulator. 3. Set the simulator to output a temperature of 34ºC. 4. Check that the displayed TEMP value on the Trio Monitor is 34ºC +/- .1ºC. 5. Change the simulator to output a value of 40ºC. 6.
3.0 Parts 3.
Exploded Views of the Trio Monitor 4 Parts 6 5 6 11 6 10 7 6 8 6 52 6 9 FIGURE 3-2 Front Panel Sub-Assembly 3-2 0070-10-0591-01 Trio™ Service Manual
Parts Exploded Views of the Trio Monitor 16 18 17 15 19 21 14 20 79 13 12 22 23 24 FIGURE 3-3 Rear Case Sub-Assembly Trio™ Service Manual 0070-10-0591-01 3-3
See Figure 3-11 32 30 See Figures 3-9 and 3-10 33 34 3-4 0070-10-0591-01 35 38 29 31 28 See Figures 3-7 and 3-8 73 27 25 42 71 43 26 Exploded Views of the Trio Monitor FIGURE 3-4 Main Chassis Sub-Assembly (SLA) Trio™ Service Manual Parts
See Figure 3-11 See Figures 3-9 and 3-10 32 69 70 38 28 72 78 Trio™ Service Manual See Figures 3-7 and 3-8 73 27 42 25 43 71 26 Parts 33 34 FIGURE 3-5 Main Chassis Sub-Assembly (Li Ion) 0070-10-0591-01 3-5 Exploded Views of the Trio Monitor
Exploded Views of the Trio Monitor Parts 46 45 47 49 48 44 44A FIGURE 3-6 Display Sub-Assembly 3-6 0070-10-0591-01 Trio™ Service Manual
Parts Exploded Views of the Trio Monitor 39 41 61 63 62 FIGURE 3-7 Masimo SPO2 Sub-Assembly Trio™ Service Manual 0070-10-0591-01 3-7
Exploded Views of the Trio Monitor Parts 41 65 39 66 67 FIGURE 3-8 Nellcor SPO2 Sub-Assembly 3-8 0070-10-0591-01 Trio™ Service Manual
Parts Exploded Views of the Trio Monitor 54 52 55 61 56 57 58 64 59 50 60 FIGURE 3-9 Right Side Panel Sub-Assembly (Masimo) Trio™ Service Manual 0070-10-0591-01 3-9
Exploded Views of the Trio Monitor Parts 54 52 65 55 56 57 58 68 59 50 60 FIGURE 3-10 Right Side Panel Sub-Assembly (Nellcor) 3 - 10 0070-10-0591-01 Trio™ Service Manual
Parts Exploded Views of the Trio Monitor 37 80 2 76 81 82 83 77 84 36 86 85 FIGURE 3-11 Rear Panel Sub-Assembly Trio™ Service Manual 0070-10-0591-01 3 - 11
Parts Listing 3.2 Parts Parts Listing REF.
Parts Trio™ Service Manual Parts Listing REF.
Parts Listing 3 - 14 Parts REF.
4.0 Repair Information 4.1 Introduction This chapter of the Service Manual provides the necessary technical information to perform repairs to the instrument. The most important prerequisites for effective troubleshooting are a thorough understanding of the instrument functions as well as an understanding of the theory of operation.
Introduction Repair Information This page intentionally left blank.
Repair Information Introduction POW ER INPUT 3 2 1 AC 1 FG N D AC 2 FA N 0119-00-0197 2 1 1 PE CPU BOARD POWER SUPPLY BOARD 0671-00-0235 250V /5A BA TT ERY BD. (LEAD ACID) BV+ PB BV- 2.54x 6 18V GN D 0012-00-1494 1 1 2.0x 2 1 3.96x 3 BV+ PB BV- 5V GN D 12V 12V GN D Vb ak 1 1 2.54x 2 12V GN D 5V BRG 0012-00-1486 12V GN D 5V BRG NC KEYPAD 0671-00-0237 or 0671-00-0064 2.54x 6 1 2.0x8 1 PC O N BCON PL ED GN D SC L SD A PC O N BCON PL ED GN D SC L SD A 2.
Introduction Repair Information CPU PC BD. POWER INPUT POWER SUPPLY BOARD 0671-00-0235 250V/5A 3 2 1 AC1 FGND AC2 FAN 0119-00-0197 PE 2 1 1 BATTERYBD. (LEAD ACID) BV+ PB BV- 2.54x6 18V GND 0012-00-1494 BV+ PB BV- 1 1 2.0x2 1 3.96x3 1 2.0x6 1 2.54x4 +18V GND 1 2.54x2 0012-00-1478 5V GND 12V 12V GND Vbak 5V GND 12V 12V GND Vbak PCON BCON PLED GND SCL SDA 0012-00-1486 12V GND 5V BRG KEYPAD 1 2.54x6 1 2.0x8 0012-00-1477 PCON BCON PLED GND SCL SDA 0671-00-0236 or 0671-00-0056 1 2.
Repair Information Introduction FIGURE 4-3 CPU PCB Connector Reference Trio™ Service Manual 0070-10-0591-01 4-5
Introduction Repair Information FIGURE 4-4 Power Supply PCB (Lead Acid Battery) Connector Reference 4-6 0070-10-0591-01 Trio™ Service Manual
Repair Information Introduction BATTERY (LI ION) P2 POWER/P12 POWER/P11 P4 P5 FAN DISPLAY REC P3 P6 P7 FIGURE 4-5 Power Supply Pcb (Lithium Ion Battery) Connector Reference Trio™ Service Manual 0070-10-0591-01 4-7
Introduction Repair Information This page intentionally left blank.
Repair Information Introduction Power Supply Bd. (Li Ion) 0671-00-0051 Power Input 250V/5A P1 3 2 1 AC1 FGND AC2 PE P6 2 1 1 17.6V GND Fan 0119-00-0197 1 P2 VBAT BC NTC GND VBAT BC NTC GND 1 Charge Ass'y BD. (Li ion) 0386-00-0303 1 P7 12V GN 5V BRG NC 12V GND 5V BRG 1 I 1 Power Supply Control Bd. NC NC GND NC VS NC VBAT NC NC 1 2 SDA SCL GND PCON 2.54x9x2 12V 5B 5V VC 17 18 17.6V 1 P5 5V GND 12V 12V BC Vbak To CPU PC BD. (P11) P4 PCON BCON PLED GND SCL SDA To CPU PC BD.
Single Temp Cable Assembly 4.2 Repair Information Single Temp Cable Assembly P/N 0012-00-1468 180 1 3 2 1 2 Temperature socket 1 red 2 black T1 AGND FIGURE 4-8 Single Temp Cable Assembly 4.
Repair Information 4.4 Troubleshooting Troubleshooting During transportation, storage and use of the Trio Monitor, various factors may result in Trio Monitor failures or affect normal operation of the device. Some of these factors include: Unstable network voltage, changing environmental temperature, physical damage or component aging. During failure conditions, qualified service technicians should perform module-level service according to the failure classification listed in the table below.
Troubleshooting Repair Information Operation, Recording MESSAGE/PROBLEM REASON SOLUTION Keys or Navigator™ knob disabled Keypad or Navigator knob is damaged Replace keypad or Navigator knob Keypad cable is damaged Replace or repair keypad cable Keypad failure Replace keypad Speaker or cable failure Replace speaker or cable Recorder has no paper or roller lever is not pressed down Install paper and press down the roller lever Recorder failure Replace the recorder Recorder cable is damaged
Repair Information Troubleshooting Parameter Failure MESSAGE/PROBLEM REASON SOLUTION No ECG waveform Poor connection of ECG electrodes Use new electrodes to ensure good contact No square waveform exists during CAL self-test Replace ECG/RESP module RL electrode is disconnected Connect RL electrode ECG waveform interference or abnormal ECG waveform No RESP waveform or abnormal RESP waveform TEMP value is incorrect HR value is inaccurate Replace ECG/RESP module Electrodes are connected incorrec
Disassembly Instructions 4.5 Repair Information Disassembly Instructions Before disassembling the unit, perform the following: • Power down the unit and remove the line cord • Remove all cable assemblies from the right side and rear of the unit • Remove any batteries that were installed • Perform all work on a properly grounded station CAUTION: 4.5.
Repair Information 4.5.5 Disassembly Instructions Removal of PCB Chassis Assembly 1. Remove four (4) 3 x 6 mm Phillips panhead screws from rear housing. 2. Remove two (2) 4 x 10 mm Phillips panhead screws from bottom of rear housing. 3. Carefully pull entire PCB chassis assembly, along with right side parameter panel, forward until it is completely removed from the rear housing. 4. Disconnect cable assemblies from connectors JP1 and JP3 of thermal printhead Recorder. 4.5.
Disassembly Instructions 4.5.10 Repair Information Removal of NIBP/IBP PCB Mounting Plate 1. Remove two (2) 3 x 6 mm Phillips panhead screws at upper right side center and upper left side center of PCB chassis assembly. 2. Remove one (1) 3 x 6 mm Phillips panhead screw at lower left of NIBP/IBP pcb mounting plate. 3. If IBP PCB is installed, remove cable assemblies from PD1 and P2 of IBP PCB. 4. Remove cable assembly from P1 of NIBP PCB. Remove polyurethane hose from Rectus fitting on parameter panel. 5.
Repair Information 4.
ECG Shielded Lead Wires 4.
Repair Information ECG Shielded Lead Wires P/N 0012-00-1261-XX - 01 18" snap 5 lead set Domestic - 14 3/40" 2/60" snap 5 lead set International - 02 24" snap 5 lead set Domestic - 07 18" snap 3 lead set Domestic - 03 40" snap 5 lead set Domestic - 08 24" snap 3 lead set Domestic - 04 18" snap 5 lead set International - 09 40" snap 3 lead set Domestic - 05 24" snap 5 lead set International - 10 18" snap 3 lead set International - 06 40" snap 5 lead set International - 11 24" snap 3 lead set Inte
Trio Wall Mounts and Rolling Stand 4.
Repair Information Trio Wall Mounts and Rolling Stand Trio VHM Wall Mount Kit 0040-00-0337-02 Trio Wall Mount Kit 0040-00-0337-01 FIGURE 4-9 Mounting bracket kits Trio™ Service Manual 0070-10-0591-01 4 - 21
Trio Wall Mounts and Rolling Stand Repair Information FIGURE 4-10 Trio roll stand Replacement Parts, Trio Rolling Stand DESCRIPTION PART NUMBER Trio rolling stand, value TRIOROLLSTD Trio monitor mounting kit 0406-00-0856-01 Casters, Non locking 0401-00-0045 Casters, Locking 0401-00-0046 Utility basket 0202-00-0166 * 4 - 22 Included in Trio Monitor Mounting Kit.
5.0 Appendix 5.1 System Alarm Prompts PROBLEM/MESSAGE REASON SOLUTION BATTERY VOLTAGE TOO LOW When battery voltage is too low, the monitor will automatically shut down within 5 minutes Use AC power supply ECG WEAK SIGNAL The patient's ECG signal is too weak Check if the electrodes and lead wires are connected correctly. Check the current condition of the patient. NO PULSE The patient's pulse signal is too weak or non-existent Check the connection of the sensor.
System Alarm Prompts Appendix PROBLEM/MESSAGE REASON SOLUTION ECG RA LEAD OFF The RA lead wire of ECG is not connected correctly Check the connection of RA lead wire ECG C LEAD OFF The C lead wire of ECG is not connected correctly Check the connection of C lead wire ECG F LEAD OFF The F lead wire of ECG is not connected correctly Check the connection of F lead wire ECG L LEAD OFF The L lead wire of ECG is not connected correctly Check the connection of L lead wire ECG R LEAD OFF The R lead
Appendix System Alarm Prompts PROBLEM/MESSAGE REASON SOLUTION *XX RANGE EXCEEDED The measured value of XX parameter has exceeded the system measurement range If message does not clear contact Customer Service. REAL CLOCK NEEDSET If the monitor displays 2000-11, this system prompt reminds the user that the current system time is inaccurate Reset the system time. After setting the time, the user should cycle power. This will avoid any time storage errors.
System Alarm Prompts Appendix PROBLEM/MESSAGE REASON SOLUTION CELL BAT TOO HIGH Cell battery is defective. Incorrect cell battery installed CELL BAT TOO LOW The cell battery has become depleted. The cell battery is not installed or the connection is loose. Replace the battery. If message does not clear contact Customer Service.
Appendix System Alarm Prompts PROBLEM/MESSAGE REASON SOLUTION LOOSE CUFF The NIBP cuff is not connected correctly Re-connect the NIBP cuff. AIR LEAK The NIBP cuff is not connected correctly or there are leaks in the airway Check the connection of each part or replace cuff and hose. If message does not clear contact Customer Service. OVER PRESSURE NIBP is in an overpressure condition. Check for occlusion in cuff and hose. Measure again. If message does not clear contact Customer Service.
0070-10-0591-01 Revision G July 17, 2010
Mindray DS USA, Inc. • 800 MacArthur Boulevard • Mahwah, NJ 07430 • USA • Dom. Customer Service: 1.800.288.2121 • Intl. Customer Service: +1.201.995.8000 • Dom. Fax: 1.800.926.4275 • Intl. Fax: +1.201.995.8680 • www.mindray.com Mindray Medical Netherlands B.V.• P.O. Box 26 • 3870 CA Hoevelaken • The Netherlands • Tel: +31 33 25 44 911 • Fax: +31 33 25 37 621 Mindray (UK) Limited • 3 Percy Road • St.