8.4.7 Enabling Pacer Rejection
The pace pulse rejection function is disabled by default. To enable this function, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG dialog.
2. Select the Pacer tab.
3. Switch on Pacer Reject.
NOTE
When pace pulses are detected, the pace pulse markers “|” are shown on the ECG waveforms. Pacer
Rejection setting has no impact on the display of pace pulse markers “|”.
You can switch on pacer rejection only when Paced is set to Yes. If Paced is set to no, the Pacer Reject
function is disabled.
8.5 Changing ECG Settings
8.5.1 Choosing an ECG Screen
When monitoring ECG, several display options are available.
For 3-lead ECG monitoring, only normal screen is available.
For 5-lead ECG monitoring, besides the normal screen, you can also choose Full-screen (7-lead) or Half
Screen (7-lead).
For 6-lead ECG monitoring, besides the normal screen, you can also choose Full-screen (8-lead) or Half
Screen (8-lead).
For 12-lead ECG monitoring, besides the normal screen, you can also choose Full-screen (7-lead), Half
Screen (7-lead), or 12-Lead (12-lead analysis).
To choose the desired screen configuration, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG dialog.
2. Select Full-Screen, Half-Screen, or 12-Lead when visible.
To return to the normal screen, press the highlighted button to unselect that screen.
Alternately you can choose the desired screen configuration as follows:
1. Select the Screen Setup quick key Choose Screen or select the Main Menu quick key from the
Display column select Choose Screen.
2. Select ECG Full-Screen, ECG Half-Screen, or ECG 12-Lead.
8.5.2 Setting ECG Alarm Properties
To select the desired ECG alarm options, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG dialog.
2. Select the Alarm tab.
3. Select the desired alarm options.
8.5.3 Changing ECG Wave Settings
8.5.3.1 Selecting the Leads of Displayed ECG Waveforms
To select the desired ECG waveforms to be displayed, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG dialog.
2. Select the Setup tab.
3. Select ECG to set the lead of each displayed ECG waveform.