7.12.2 Exiting the CPB Mode
To exit the CPB mode, choose either of the following ways:
Select the CPB Mode quick key.
Select the Main Menu quick key from the Alarm column select Exit CPB Mode.
7.13 Intubation Mode
Intubation mode is available for Resp, CO
2
, and AG monitoring. When performing intubation during general
anesthesia, you can put the monitor in the intubation mode in order to inactivate unnecessary alarms.
In the intubation mode, Resp, CO
2
and AG related physiological alarms are switched off.
7.13.1 Entering the Intubation Mode
To enter the intubation mode, choose either of the following ways:
Select the Intubation Mode quick key.
From the bottom of the Resp, CO2, or AG dialog, select Intubation Mode.
Select the Main Menu quick key from the Alarm column select Intubation Mode.
7.13.2 Exiting the Intubation Mode
To exit the intubation mode, choose either of the following ways:
Select the Intubation Mode quick key.
From the bottom of the Resp, CO2, or AG dialog, select Exit Intubation Mode.
Select the Main Menu quick key from the Alarm column select Exit Intubation Mode.
7.14 Testing Alarms
The monitor automatically performs a selftest at startup. Check that an alarm tone is heard, the alarm lamp
illuminates, one after the other, in red, yellow, and cyan. This indicates that the visible and audible alarm
indicators functions correctly.