7.10 Nurse Call
The monitor provides a nurse call connector to output a nurse call signal when a user-defined alarm occurs. To
obtain a nurse call signal, use the nurse call cable to connect the hospital nurse call system with the monitors
nurse call connector.
Alarms are indicated on the nurse call device only when the following conditions are met:
The nurse call system is enabled.
A user-defined alarm occurs.
Alarms are not paused or reset.
WARNING
Do not rely exclusively on the nurse call system for alarm notification. Remember that the most
reliable alarm notification combines audible and visual alarm indications with the patient’s clinical
condition.
7.11 Calling for Help
This monitor can be configured to call monitors in the same department, the central station, and the nurse call
system so that nearby doctors and nurses can come for help.
7.11.1 Issuing the Call Help Signal
To call help, select the Call Help quick key and select OK from the popup dialog box. If OK is not selected within
5 seconds then the monitor will automatically send out the call help signal.
After the call help signal is sent out, the Call Help quick key flashes in red. Select the Call Help quick key again to
stop the call help signal.
Monitors receiving the call help signal issue a sound and a dialog box pops up indicating which monitor is
calling. Select OK to acknowledge the call and stop the sound at this monitor.
NOTE
The call help function works only when the monitor is connected to the network.
The call help sound may disturb patients in the same department.
7.12 CPB Mode
The CPB (Cardiopulmonary Bypass) mode is activated only if you set the department to OR.
In the CPB mode, all the physiological alarms and technical alarms are switched off. So when performing CPB,
you can put the monitor in the CPB mode to inactivate unnecessary alarms.
7.12.1 Entering the CPB Mode
To enter the CPB mode, choose either of the following ways:
Select the CPB Mode quick key.
Select the Main Menu quick key from the Alarm column select CPB Mode.
In the CPB mode, CPB Mode is displayed in the physiological alarm area with a red background color.
NOTE
When the CPB mode is entered, the monitor stops all NIBP measurements. You can restart NIBP
measurements after entering the CPB mode.