7.6.5 Setting the Apnea Delay Time
To set the zero respiration rate delay time, follow this procedure:
1. Access the Setup page in either of the following ways in either of the following ways:
Select the Alarm Setup quick key select the Setup tab.
Select the Main Menu quick key from the Alarm column select Setup.
IBP: CVP/
LAP/
RAP/
UVP/P1-
P4
(Venous
pressure)
IBP-M
(mmHg)
Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40
CO
2
EtCO
2
(mmHg)
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement range
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
+ 6)
36 to 45:
(etCO
2
+ 6)
Same as the
measurement range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
AG EtCO
2
Same as the CO
2
module
FiCO
2
Same as the CO
2
module
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
awRR+25 or
85 (whichever
is smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
FiAA/EtAA Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiO
2
/EtCO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiN
2
O/
EtN
2
O
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
C.O. TB (°C) Adult: (TB - 1)
Pediatric: N/A
N/A Adult: (TB + 1)
Pediatric: N/A
N/A Same as the
measurement range
C.I. N/A N/A N/A N/A N/A
TFC N/A N/A N/A N/A N/A
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate