User's Manual

Table Of Contents
7 - 4
7.6 Changing Alarm Settings
Alarms settings can be changed by selecting Main Menu and choosing from the list in the Alarm column.
7.6.1 Setting Parameter Alarm Properties
To set parameter alarm properties, follow this procedure:
1. Access the Limits page in either of the following WAYS:
Select the Alarm Setup quick key.
Select the Main Menu quick key from the Alarm column select Limits.
2. Select a parameter tab and set alarm properties as desired.
You can also change the alarm properties of individual parameter from the corresponding parameter dialog.
NOTE
The monitor may be configured to require an password or login to change alarm settings. For more
information, refer to 25.13.3Selecting Password for User Authentication.
7.6.2 Setting Alarm Tone Properties
7.6.2.1 Changing the Alarm Volume
To change the alarm volume, follow this procedure:
1. Access the Setup page in either of the following ways:
Select the Alarm Setup quick key select the Setup tab.
Select the Main Menu quick key from the Alarm column select Setup.
2. Set Alarm Volume. The alarm volume can be set from X to 10, in which X is the minimum volume allowed
by your department, and 10 is the maximum volume.
3. Select High Alarm Volume to set the volume of the high priority alarm.
4. Select Reminder Volume to set the volume of the reminder tone.
NOTE
Alarm volume can only be set to 0 if the monitor is connected to a Central Monitoring System.
When the alarm volume is set to 0, the alarm sound is turned off and the audio off symbol appears
on the screen.