User's Manual

Table Of Contents
5 - 2
NOTE
The monitor will reload the configuration if you changed the patient category.
5.2.3 Loading Patient Information from the CMS
If the monitor is connected to the central monitoring system (CMS). You can load patient information from the
CMS to the monitor. To do so, follow this procedure:
1. Enter the Find Patient dialog in either of the following ways:
Select the Main Menu quick key from the Patient Management column select Find Patient.
From the Patient Management dialog select Find Patient.
2. Input query criteria.
3. Select Search. Then a list pops up, including all the patients that meet the query criteria.
4. Select a patient from the patient list, and then select Admit. Corresponding patient information in the
monitor will be updated.
NOTE
If ADT Query is switched on, input query criteria from the Discharged Patient page.
5.2.4 Loading Patient Information from the ADT Server
If the monitor is connected with the Admit-Discharge-Transfer (ADT) server through the eGateway. You can load
patient information from ADT server to the monitor.
To do so, follow this procedure:
1. Enter the Find Patient dialog in either of the following ways:
Select the Main Menu quick key from the Patient Management column select Find Patient.
Select Find Patient from the Patient Management dialog.
2. Input query criteria.
3. Select Search. Then a list pops up, including all the patients that meet the query criteria.
4. Select a patient from the patient list, and then select Admit. Corresponding patient information in the
monitor will be updated.
NOTE
You can load patient information from the ADT server only when ADT Query is enabled. For more
information, see 25.12.12Using the ADT Gateway.
The monitor can automatically upload the patient information from the ADT server when:
The configured keywords of the monitor and the eGateway are the same.
The IP address of the monitor is added to the eGateway configuration dialog.
Loading patient information from the ADT server updates only patient information in the monitor.
The patient’s monitoring data is not changed and the patient is not discharged.
5.3 Exporting Patient Data
To export the data of the current patient and discharged patients, follow this procedure:
1. Connect the USB drive to the monitor’s USB connector.
2. Access the Discharged Patients dialog box by either of the following ways:
Select the Discharged Patients quick key.
Select the Main Menu quick key from the Patient Management column select Discharged
Patients.
3. From the patient list select desired patients.