4.7.1 Entering the Remote View Screen
To enter the Remote View screen, choose one of the following ways:
Select the Remote View quick key.
Select the bed at the alarm watch tile on the main screen. For more information, see 4.7.10.2Displaying the
Alarm Watch Tile on the Main Screen.
Select the Screen Setup quick key select the Primary Display tab or Secondary Display tab (as
available) select the Choose Screen tab select Remote View.
4.7.2 About the Remote View
The following figure shows the Remote View window.
(1)
Alarm watch area
Display all the monitored remote beds.
Each bed displays the room number, bed number, connection status and alarm status. The
background color indicates the alarm status on the corresponding bed.
Background Color Description
Green No alarm is occurring to the bed.
Red The remote device is disconnected or a high priority alarm is occurring. The high
priority alarm currently is the highest alarm level on the bed. If the remote device is
disconnected, the icon is displayed.
Yellow The medium priority alarm is occurring. The medium priority alarm currently is the
highest alarm level on the bed.
Cyan The low priority alarm is occurring. The low priority alarm currently is the highest
alarm level on the bed.
Grey The bed is in the standby mode.
(2)
Main body
Display the patient’s information, alarm status and messages, waveforms, measurements, etc. of the
selected bed. This bed is called main bed.