User's Manual

Table Of Contents
4 - 6
4.5.4 Setting OxyCRG Parameters
Select parameter trends or compressed waveform to set parameters and the compressed waveform you want to
display. The selected parameters will be used for ABD event calculation.
4.5.5 Setting the Threshold of ABD Events
Select any parameter trend or the compressed waveform to perform the following setup:
Set the threshold of ABD events.
Set Event Storage Format:
1 min+3 min: stores data one minute before and three minutes after the event.
3 min+1 min: stores data three minutes before and one minute after the event.
2 min+2 min: stores data two minutes before and two minutes after the event.
The stored data includes the trends of the OxyCRG parameters, compressed waveform, alarm thresholds, NIBP,
and Temp measurements.
4.5.6 Editing ABD Events
To edit ABD events, follow this procedure:
1. Select the Mark button to enter the Mark dialog box.
2. Drag the event list upwards and downwards to select the desired event.
3. Select the patients status when the event happens.
4. Select Save.
4.6 The SpO
2
Screen
For neonatal patients, if you are only concerned with the patient’s SpO
2
and pulse rate, you can use the SpO2
screen.
The SpO
2
screen displays SpO
2
related data. It also displays realtime Temp and NIBP measurements.
NOTE
The SpO
2
screen is intended for neonatal patient only.
4.6.1 Entering the SpO
2
Screen
To enter the SpO
2
screen, choose any of the following ways:
Swipe left or right on the touchscreen with two fingers to switch to the SpO
2
screen.
Select the Screen Setup quick key select the Choose Screen tab select SpO2 Screen.
Select the Main Menu quick key from the Display column select Choose Screen select SpO
2
Screen.