D.2.16 Other System Technical Alarm Messages
Alarm message Default priority Indication on
alarm reset
Cause and solution
Storage Card Error High C The storage card fails or files are damaged.
Restart the monitor to format the storage card. If
the alarm persists, contact your service
personnel.
Loading Default Config
Failed
Low A The default configuration is not correctly loaded.
The monitor will restore to the factory default
configuration for the current patient category.
XX Conflicts
(XX refers to the module
label)
Prompt / The same type of corresponding module being
used exceeds the supported number. Remove
the conflict module.
XXX Measurement has
been closed
(XX refers to the module
label)
Prompt / The parameter module is disabled. Switch on the
module if you want to use it. For more
information, see
3.11.1Switching On or Off a
Parameter
.
The display setup for XXX
is disabled.
(XX refers to the
parameter label)
Prompt / The parameter of the newly inserted module is
not displayed on the screen. Select a desired area
to display the parameter numerics and
waveforms. For more information, see
25.11The
Other Settings
.
The patient data storage
space is nearly full. Please
delete some discharged
patients.
Configurable B Delete unnecessary earlier discharged patient.