D.2.5 Temp Technical Alarm Messages
Note: XX represents a temperature site, for example skin, core, axil, T1, and so on.
D.2.6 NIBP Technical Alarm Messages
SpO2 No Pulse Low C The SpO
2
sensor failed to obtain pulse signal.
Check the patient’s condition and replace the
sensor application site. If the alarm persists,
replace the sensor.
SpO2 Sensor
Incompatible
Low C Incompatible or an unspecified SpO
2
sensor is
used. Use specified sensors.
SpO2 Low Signal Quality Low C 1. Check the sensor and sensor position.
2. Make sure the patient is not shivering or
moving.
3. The patient’s pulse may be too low to be
measured.
SpO2 Interference Low C The SpO
2
signal has been interfered. Check for
any possible sources of signal noise and check
the patient for excessive motion.
SpO2 Sensor Error Low C Replace the sensor and measure again.
SpO2 Searching Pulse Prompt / SpO
2
is searching for pulse.
SpO2 Low Perfusion Prompt / The SpO
2
sensor is not properly placed or the
patient’s perfusion index is too low.
1. Check the sensor and sensor position.
2. Reposition the sensor if necessary.
Alarm message Default priority Indication on
alarm reset
Cause and solution
Alarm message Default priority Indication on
alarm reset
Cause and solution
T XX Sensor Off Low A Check the sensor connection and reconnect the
sensor.
Alarm message Default priority Indication on
alarm reset
Cause and solution
NIBP Cuff Loose Low A There is a leak in the cuff or air tubing. Use a cuff
of correct type based on the patient size. Apply
the cuff and connect the air tubing as instructed
in the manual.
NIBP Cuff or Airway Leak Low A Check the NIBP cuff and pump for leakages.
NIBP Airway Error Low A The air tubing may be occluded. Check the air
tubing for an occlusion or kinking. If the alarm
persists, contact your service personnel.
NIBP Weak Signal Low A The patient’s pulse is weak or the cuff is loose.
Check the patient’s condition and replace the cuff
application site.
NIBP Overrange Low A The measured NIBP value exceeds the module
measurement range. Check the patient’s
condition.
NIBP Excessive Motion Low A Check the patient’s condition and reduce patient
motion.