User's Manual

Table Of Contents
C - 11
Art-M Extreme Alarm switch Off
High limit Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit Adult: 55 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority High
Alarm Outputs Off
CPP Alarm switch On
High limit Adult: 130 mmHg
Pediatric: 100 mmHg
Neonate: 90 mmHg
Low limit Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority Med
Alarm Outputs Off
Measure (for P1, P2) All
Measure (for P3, P4) Mean only
Sensitivity Med
Speed 25 mm/sec
Scale (mmHg) ICP/RAP/LAP/UVP
venous pressure
0-20
Art/Ao/BAP/FAP/LV/P1/
P2 arterial pressure
0-160
UAP/P3/P4 venous
pressure
0-80
PA/CVP 0-30
PPV Measure Off
PPV Source Auto
PAWP Reference Waveform 1 II
Reference Waveform 2 Resp
Speed 12.5 mm/sec
PA Scale (mmHg) 0-30
Overlapping
Waveform Setup
Left Scale (mmHg) 0-160
Right Scale (mmHg) 0-20
CVP Scale (mmHg) 0-30
ICP Scale (mmHg) 0-20
PA Scale (mmHg) 0-30
Speed 25 mm/sec
Gridlines Off
Display Format Sys/Dia (Mean)
Item Default Setting