NIBP-M Extreme Alarm switch
ICU, OR, CCU: On
Other departments: Off
High limit Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit Adult: 45 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority High
Alarm Outputs Off
Initial Pressure Adult: 160 mmHg
Pediatric: 140 mmHg
Neonate: 90 mmHg
Interval
OR: 5 min
NICU: 30 min
Other departments: 15 min
Start Mode Clock
NIBP End Tone Off
Venipuncture Pressure Auto
Display Format Sys/Dia (Mean)
Nighttime Start Time 22:00
End 07:00
Alarm Limit Display
ICU, OR, CCU: Off
Other departments: On