User's Manual

Table Of Contents
C - 2
C.1.2 Arrhythmia Default Settings
C.1.2.1 Arrhythmia Alarm Default Settings
Item Alarm Switch Priority Alarm Outputs
Asystole On High, unadjustable Off
V-Fib/V-Tach On High, unadjustable Off
V-Tach On High, unadjustable Off
Vent Brady On High, unadjustable Off
Extreme Tachy On High, unadjustable Off
Extreme Brady On High, unadjustable Off
R on T
CCU: On
Other departments: Off
Med Off
Run PVCs Off Low Off
Couplet Off Prompt Off
Multiform PVC Off Med Off
PVC Off Prompt Off
Bigeminy
CCU: On
Other departments: Off
Med Off
Trigeminy
CCU: On
Other departments: Off
Med Off
Tachy Off Med Off
Brady Off Med Off
Pacer Not Capture Off Prompt Off
Pacer Not Pacing Off Prompt Off
Missed Beats Off Prompt Off
Nonsus V-Tach
CCU: On
Other departments: Off
Med Off
Vent Rhythm
CCU: On
Other departments: Off
Med Off
Pause Off Low Off
Irr. Rhythm Off Prompt Off
A-Fib Off Prompt Off
PVCs/min
CCU: On
Other departments: Off
Med Off
Pauses/min
CCU: On
Other departments: Off
Med Off