*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2)in
terms of mean error and standard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
standard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
A.13.7 IBP Specifications
Accuracy Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure range
(mmHg)
Adult: 80 to 280
Pediatric: 80 to 210
Neonate: 60 to 140
Default initial cuff inflation pressure
(mmHg)
Adult: 160
Pediatric: 140
Neonate: 90
Software overpressure protection Adult: 297±3 mmHg
Pediatric: 297±3 mmHg
Neonate: 147±3 mmHg
Static pressure measurement range 0 mmHg to 300 mmHg
Static pressure measurement accuracy ±3 mmHg
Recovery time <15 s (after defibrillation)
Alarm limit Range (mmHg)
Step (mmHg)
NIBP-S High Adult: (low limit + 5) to 290
Pediatric: (low limit + 5) to 240
Neonate: (low limit + 5) to 140
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Low 25 to (high limit - 5)
NIBP-M High Adult: (low limit + 5) to 260
Pediatric: (low limit + 5) to 215
Neonate: (low limit + 5) to 125
NIBP-M Low 15 to (high limit - 5)
NIBP-D High Adult: (low limit + 5) to 250
Pediatric: (low limit + 5) to 200
Neonate: (low limit + 5) to 115
NIBP-D Low 10 to (high limit - 5)
Standard Meet the standard of IEC 60601-2-34: 2011
Technique Direct invasive measurement
IBP
Measurement range -50 to 360 mmHg
Resolution 1 mmHg
Accuracy ±2% or ±1 mmHg, whichever is greater (excluding sensor error)
Refreshing rate ≤1 s