A.13.4 PR Specifications
PR from Masimo SpO
2
PR from Mindray SpO
2
Module
PR from Nellcor SpO
2
Module
PR from NIBP Module
Response time ≤30 s (normal perfusion, no disturbance, SpO2 value sudden change from
70% to 100%)
Recovery time <15 s (after defibrillation)
Accuracy 70 to 100%: ±2% (adult/pediatric)
70 to 100%: ±3% (neonate)
0% to 69%: Not specified.
When the SpO
2
sensor is applied for neonatal patients as indicated, the specified accuracy range is increased by ±1%, to
compensate for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood.
Alarm limit Range Step
PR High PR≤40bpm: (low limit + 2 bpm) to 40 bpm
PR > 40 bpm: (low limit + 5 bpm) to 295 bpm
PR≤40: 1
PR>40: 5
PR Low PR≤40bpm: 16 bpm to (high limit - 2 bpm)
PR > 40 bpm: 40 bpm to (high limit - 5 bpm)
Measurement range 25 to 240 bpm
Resolution 1 bpm
Response time ≤20 s (with normal perfusion, no disturbance, and a PR value transition from
25 to 220 bpm)
Accuracy ±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refresh rate ≤1 s
Sensitivity High, Medium, Low
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time <30 s (normal perfusion, no disturbance, PR value sudden changes from 25
to 220bpm)
Accuracy ±3 bpm
Refreshing rate ≤1 s
Sensitivity High, Medium, Low
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time ≤30 s (normal perfusion, no disturbance, PR value sudden change from 25
to 250 bpm)
Accuracy 20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate ≤1 s