A.4 Power Supply Specifications
A.4.1 External Power Supply Specifications
A.4.2 Battery Specifications
AC Power
Input voltage 100 to 240 VAC (±10%)
Input current 2.0 to 0.9 A
Frequency 50/60 Hz (± 3 Hz)
Battery LI13I001A Type Rechargeable lithium-Ion battery (non-smart battery)
Voltage 10.95V
Capacity 2600 mAh
Battery LI23S002A Type Rechargeable lithium-Ion battery (non-smart battery)
Voltage 10.95V
Capacity 4500 mAh
Battery LI23I003A Type Smart rechargeable lithium-Ion battery (smart
battery)
Voltage 10.8V
Capacity 5600 mAh
Maximum number of batteries configured ePM 10M: only one battery can be connected.
ePM 12M: at most two batteries can be connected at the same
time, including one battery built inside, the other one
connected from an external battery compartment.
ePM 15M: at most two batteries can be connected at the same
time.
Run time Model ePM 12M/ePM 15M ePM 10M
Battery LI13I001A
(one battery
connected)
≥ 2 hours ≥ 2 hours
Battery LI23S002A
(one battery
connected)
≥ 4 hours ≥ 4 hours
Battery LI23I003A
(one battery
connected)
≥ 4.5 hours ≥ 6 hours
Battery LI23I003A
(two batteries
connected)
≥ 9 hours /
when the monitor is powered by a new fully-charged battery at 25 ºC±5 ºC
with 5-lead ECG and SpO
2
cable connected, auto NIBP measurements at an
interval of 15 minutes, and screen brightness set to 1.
Shutdown delay: at least 15 minutes after the low battery alarm first occurs.