The environmental specification of unspecified parameter modules are the same as those of the
main unit.
Main Unit
Item Temperature (ºC) Relative humidity
(noncondensing) (%)
Barometric
Operating Condition
0 to 40
(0 to 35 configured with
AG module)
15 to 95 427.5 to 805.5 mmHg
(57 to 107.4 kPa)
Storage Condition -20 to 60 10 to 95 120 to 805.5 mmHg
(16 to 107.4 kPa)
Microstream CO
2
Module
Item Temperature (ºC) Relative humidity
(noncondensing) (%)
Barometric
Operating Condition
0 to 40
15 to 95 430 to 790 mmHg
(57.3 to 105.3 kPa)
Storage Condition -20 to 60 10 to 95 430 to 790 mmHg
(57.3 to 105.3 kPa)
Sidestream CO
2
Module
Item Temperature (ºC) Relative humidity
(noncondensing) (%)
Barometric
Operating Condition
5 to 40 15 to 95 430 to 790 mmHg
(57.3 to 105.3 kPa)
Storage Condition -20 to 60 10 to 95 430 to 790 mmHg
(57.3 to 105.3 kPa)
Mainstream CO
2
Module
Item Temperature (ºC) Relative humidity
(noncondensing) (%)
Barometric (mmHg)
Operating Condition
0 to 40
10 to 90 430 to 790 mmHg
(57.3 to 105.3 kPa)
Storage Condition -20 to 60 10 to 90 430 to 790 mmHg
(57.3 to 105.3 kPa)
AG Module
Item Temperature (ºC) Relative humidity
(noncondensing) (%)
Barometric
Operating Condition
10 to 40 15 to 95 525 to 805.5 mmHg
(70 to 107.4 kPa)
Storage Condition -20 to 60 10 to 95 525 to 805.5 mmHg
(70 to 107.4 kPa)