User's Manual

Table Of Contents
3 - 10
3.7.2.1 Entering Privacy Mode
To enter Privacy mode, choose either of the following ways:
Select the Privacy Mode quick key select OK.
Select the Main Menu quick key from the Display column select Privacy Mode select OK.
The monitor has the following features after entering Privacy mode:
The screen turns blank.
Except for the low battery alarm, the monitor inactivates alarm tones and alarm lights for all other alarms.
The monitor suppresses all system sounds, including heart beat tone, pulse tone, and prompt tone.
WARNING
.
In Privacy mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor.
Alarms are presented only at the CMS. Pay attention to potential risk.
NOTE
Privacy mode is not available if the Department is set to OR.
You cannot enter Privacy mode if a low battery alarm is active.
3.7.2.2 Exiting Privacy Mode
The monitor automatically exit Privacy mode in any of the following situations:
The monitor disconnects from the CMS.
The low battery alarm occurs.
You can also operate the touchscreen, mouse, or keyboard to manually exit Privacy mode.
3.7.3 Night Mode
Night mode is a special clinical monitoring mode used to avoid disturbing the patient. Night mode reduces
screen brightness, sound volumes, and pauses NIBP measurements.
3.7.3.1 Entering Night Mode
To enter Night mode, follow this procedure:
1. Select the Night Mode quick key, or select the Main Menu quick key from the Display column select
Night Mode.
2. Change Night mode settings if necessary.
3. Select Enter Night Mode.
Night mode settings are as follows by default:
Brightness: 1
Alarm Volume: 2
QRS Volume: 1
Key Volume: 0
NIBP End Tone: Off
Stop NIBP: Off
CAUTION
Verify Night mode settings before entering Night mode. Pay attention to the potential risk if the
setting value is low.