3. Select Add Central Station, and then input the CMS name and IP address of this department.
25.12.9 Setting Multicast Parameters
Multicast helps device discovery between monitors and other devices on the network. Devices in the same
multicast group can be mutually discovered.
To set multicast parameters, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Network Setup tabDevice Discover tab.
3. Set
Multicast TTL.
4. Set
Multicast Address.
25.12.10 Setting the CMS IP or Gateway Address
To set CMS address and the DNS name, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Network Setup tab Device Discover tab.
3. Set
Central Station Address to input the DNS name or IP address.
The Center Station Address is 0.0.0.0 by default. That is to say that the monitor can be admitted by
any CMS.
If Center Station Address is not 0.0.0.0, the monitor can only be admitted by the specified CMS. The
monitor will send unicast presence message to that CMS. If the CMS cannot discover the monitor via
multicast, it can discover the monitor via unicast.
If the monitor is connected to the CMS, the displayed IP of
Connected Central Station is the IP of this CMS.
NOTE
If there is no central station for use with this monitor, then it can connect directly to the eGateway by
entering the eGateway Address in the Central Station Address Field.
25.12.11 Setting the Network Service Quality Level
To set the quality of service (QoS), follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Network Setup tabQoS tab.
3. Select
QoS Level For Realtime Monitoring to set the service quality of network connection for realtime
monitoring, for example parameter measurements and waveforms, alarms, etc.
4. Select
QoS Level For Others to set the service quality of network connection for non-realtime monitoring,
for example history data, printing, etc.
25.12.12 Using the ADT Gateway
The ADT (admit-discharge-transfer) gateway is normally deployed in the eGateway. You can obtain patient
information from the hospital ADT server through the ADT gateway.
To configure the ADT gateway, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Network Setup tab ADT tab.
3. Select
Server Address to input the host name or IP address of the ADT gateway.
4. Select
Port to input the port of the ADT gateway.