User's Manual

Table Of Contents
25 - 13
1. Select the Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab ECG tab.
3. From the bottom left corner of the dialog select
Calibrate.
25.5.5 Setting Parameter Measurement Timeout
NIBP measurements become outline fonts after a preset time. This feature prevents older values being
misinterpreted as current measurements. To adjust the timeout period for PAWP, C.O., and NIBP, follow this
procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab select Other.
3. Set the timeout period for PAWP, C.O., and NIBP.
25.5.6 Setting Flow Rate (For Sidestream CO
2
Module Without O
2
)
When you are using the sidestream CO
2
module without the O
2
monitoring function to monitor a neonatal
patient, you can select flow rate. To do so, follow this procedure:
1. Ensure that the sidestream module without O
2
is inserted in the monitor.
2. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
3. Select the
Module tab Other tab.
4. Set
CO2 Flow Rate For Neo.
25.6 Hiding Undesired Review Tabs
If you do not need to review some items, you can hide them. To do so, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Review tab.
3. From the
Tabs page, Event page or Arrh Mark page deselect those not need to beavailable to clinicians
during review.
25.7 Setting the Printer
25.7.1 Setting the Printer Properties
To set a network printer, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Print tab.
3. In the
Printer tab, select Connection Type to choose whether you want to output patient reports via the
print server or a network printer.
4. If you set
Connection Type to Print Server, perform the following settings:
Print Server Address: select to input the Print Server name or IP address.
Port: selects to input the Print Server port. If the CMS is used as the Print Server, set Port to 6603.
Printer: select the default printer.
Resolution: select the resolution of the default printer.
5. If you set
Connection Type to Printer, perform the following settings:
Select Printer IP Address. Select to input the IP address of the desired network printer.