25.4.3 Defining the Hours of Nighttime for the ECG 24h Summary
To define the nighttime for heart rate statistics, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
CAA tab ECG 24h Summary tab.
3. Select
From and To to define the beginning and end of Nighttime.
25.5 Changing Module Settings
25.5.1 Selecting the ECG Standard
Select the ECG standard according to the leadwires you are using. To select the ECG standard, follow this
procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab ECG tab.
3. Set
ECG Standard to AHA or IEC.
25.5.2 Changing the QTc Formula
The monitor uses the Hodges correction formula by default to correct the QT interval for heart rate. To select the
QTc formula, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab ECG tab.
3. Set
QTc Formula.
Hodges:
Bazett:
Fridericia:
Framingham:
25.5.3 Setting the 12-Lead Order Switch
To set whether to send the order number of 12-lead interpretation report to the hospital information system
while saving the report, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Module tab.
3. Select the
ECG tab.
4. Set
12-Lead Order.
25.5.4 Calibrating ECG
The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG waveform
amplitude becomes greater or smaller. In that case, you need to calibrate the ECG module. To do so, follow this
procedure:
QTc QT
1.75
HeartRate
60()×+=
QTc QT
HeartRate
60
-----------------------------


1
2
-- -
×
=
QTc QT
HeartRate
60
-----------------------------


1
3
---
×=
QTc QT 154 1
60
HeartRate
-----------------------------


×
+=