User's Manual

Table Of Contents
25 - 11
The arrhythmia alarm timeout period is only applied to the medium priority chains and atrial
fibrillation chain. For the alarms in the high priority chain, alarm tone and alarm light are presented
as soon as the alarm condition is detected. Refer to 8.6.5.1Arrhythmia Alarm Chains for more
information on the arrhythmia alarm chain.
Alarm indication rules for alarms in the atrial fibrillation chain are the same with those for the
medium priority chains.
25.4 Changing CAA Settings
25.4.1 Changing EWS Settings
25.4.1.1 Displaying and Managing the Operator ID
To display the operator ID on the EWS screen, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
CAA tab select the EWS tab.
3. Switch on
Operator ID.
4. Select
Operator ID Timeout to set how long the operator ID will remain valid.
NOTE
The EWS tab is only available if a license for EWS is installed in the monitor.
25.4.1.2 Selecting the Default Scoring Tool
To set the default scoring tool for different patient catalog, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
CAA tab select EWS tab.
3. From the
Select Default Score area, set Default Adult Score, Default Ped Score, and Default Neo Score.
25.4.1.3 Managing the Scoring Tools
To manage the scoring tool, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
CAA tab select EWS tab.
3. Select the
Manage Score button.
Delete local scoring tools: from the Local page, delete any unnecessary scoring tools.
Import the desired scoring tools to the monitor: from the USB drive page, select the scoring tools on
the USB drive, and then select
Import.
NOTE
The monitor provide MEWS and NEWS by default. You cannot delete them.
25.4.2 Setting GCS Threshold for Each Consciousness Level
You can configure the threshold and color of each consciousness level. To do so, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
CAA tab select the GCS tab.
3. Set high limit, low limit for each level.