User's Manual

Table Of Contents
25 - 9
3. Set SpO2 Desat Alarm Off.
Disable: the SpO
2
Desat alarm is always on. You cannot switch it off.
Enable: you can switch off the SpO
2
Desat alarm. For more information, see 11.6.1Changing the SpO
2
Alarm Settings
.
WARNING
If a clinician switches off the SpO2 Desat alarm, the monitor will not alarm when the patient’s SpO
2
is
extremely low. This may result in a hazard to the patient.
25.3.7.7 Setting the Switch of the Apnea Alarm Off
You can choose whether switching off the apnea alarm is permissible or not. To do so, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab select the Other tab.
3. Set
Apnea Alarm Off.
Disable: the zero RR alarm is always on. The clinician cannot switch it off.
Enable: the clinician can switch off the zero RR alarm. For more information, see 10.5.1Setting the Resp
Alarm Properties.
WARNING
If a clinician switches off the zero respiration rate alarm, the monitor will not issue the zero
respiration rate alarm when that situation occurs. This may result in a hazard to the patient. Keep the
patient under close surveillance.
25.3.8 Changing Alarm Delay Settings
25.3.8.1 Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
To set the alarm delay time, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Other tab.
3. Set
Alarm Delay.
The setting of
Alarm Delay is not applied to the zero respiration rate alarms and the ST alarms. You can set Zero
RR Delay and ST Alarm Delay separately.
25.3.8.2 Setting the ST Alarm Delay Time
1. Select the Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Other tab.
3. Set
ST Alarm Delay.