25 - 7
25.3.5.5 Setting the Tone Pattern for Alarms from Remote Devices
The monitor provides the same alarm tone pattern for the remote device alarms as those for your monitor
alarms. For more information on alarm tone pattern, see
7.3.3Alarm Indicators.
To set the alarm tone pattern for the remote device alarms, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab select the Remote View tab.
3. Set
Alarm Sound to ISO, Mode 1 or Mode 2.
25.3.5.6 Switching Off the Remote Device Disconnection Alarm
The monitor can provide an alarm if remote devices, for example, a bedside monitor or a telemetry, are
disconnected. By default, the function is enabled. To disable the alarm, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Remote View tab.
Switch off
Remote Disconnected Alarm.
25.3.6 Setting Nurse Call Properties
25.3.6.1 Selecting Alarms to be Sent to the Nurse Call System
To set the type and priority of alarms that are sent to the nurse call system, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Nurse Call tab.
3. Select
Signal Type
Pulse
: the nurse call signal is a pulse signal and each pulse lasts one second. When multiple alarms
simultaneously occur, only one pulse signal is outputted. If an alarm occurs but the previous one is
not cleared, a new pulse signal will also be outputted.
Continuous: the nurse call signal lasts until the alarm ends. That is to say the duration of a nurse call
signal is equal to that of the alarm condition.
4. Select Contact Type to set the work mode of the nurse call relay.
5. Select
Alarm Priority to set the priority of alarms sent to the nurse call system.
6. Select
Alarm Type to set the type of alarms sent to the nurse call system.
25.3.6.2 Disabling the Receiving Call Help Function
If a monitor in the same department calls for help, you can receive the calling signal. This function is enabled by
default. To disable this function, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Nurse Call tab.
3. Switch off
Receive Call Help.