User's Manual

Table Of Contents
21 - 2
21.3 Stopping Recordings
Recordings can be stopped manually or automatically.
21.3.1 Stopping Recordings Manually
To manually stop a recording, press the hardkey while the recorder is creating a strip.
21.3.2 Stopping Recordings Automatically
Recordings stop automatically in the following conditions:
The recording is completed.
The recorder runs out of paper.
The recorder has an alarm condition.
The recorder module is removed from the monitor.
The monitor is turned off.
21.4 Recording Related Flags
You can find the following flags on the recording reports:
For automatically stopped recordings, there are two columns of asterisks “*” at the end of the report.
For manually or abnormally stopped recordings, there is one column of asterisks “*” at the end of the
report.
If the parameter data is from external devices connected to the monitor via the BeneLink module, the
parameter label is prefixed with the plus sign “+”.
21.5 Setting up the Recorder
To configure the recorder for waveforms and printing intervals, follow this procedure:
1. Select the
Main Menu quick key from the Report column select Record Setup.
2. In the
Record Setup dialog, select the desired waveform for Waveform 1, Waveform 2 and Waveform 3 in
turn. The recorder can record up to 3 waveforms at a time.
3. Switch on or off
IBP Overlap to enable or disable IBP recordings in the overlapping format.
When the IBP Overlap is enabled: If two or more waveforms in the selected waveforms for recording
are IBP waveforms, the IBP waveforms will be recorded in the overlapping format.
When the IBP Overlap is disabled: IBP waveforms will be recorded without overlap.
4. Select
Length to set the duration of real-time recording.
5. Select
Interval to set the time interval for automatic recording.
6. Select
Paper Speed to set the speed for recording waveforms.
21.6 Clearing Recording Tasks
To clear recording tasks, follow this procedure:
1. Select the
Main Menu quick key from the Report column select Record Setup.
2. In the
Record Setup dialog, select Clear All Tasks. This clears all queued recording tasks and stops the
current recording.