User's Manual

Table Of Contents
18 - 9
18.2.10.1 Entering the OxyCRG Review Page
Choose one of the following methods to enter the OxyCRG review page:
From the OxyCRG screen, select the ABD events list area.
Select the Review quick key select the OxyCRG tab.
Select the Main Menu quick key from the Review column select OxyCRG.
18.2.10.2 The Display of the OxyCRG Review Page
The following figure shows the OxyCRG screen:
18.2.10.3 Changing the Resolution of Trend Curves
To set the resolution of trend curves, follow this procedure:
1. Enter the OxyCRG review page.
2. Set
Zoom.
NOTE
If the Zoom is set for a large duration (for example, 8 hrs), the screen may show "No Patient Data" for
the first several minutes of data collection (for example, 6 minutes), but data will be shown after
enough data has been collected.
18.2.10.4 Printing an OxyCRG Review Report
To print an OxyCRG review report, follow this procedure:
1. Enter the OxyCRG review page.
2. Select to enter the
Print Setup dialog.
3. Select
Print. The report duration corresponds to the duration of data shown on the screen.
(3)
(4)
(2)
(2)
(1)
(1) Event title area: displays information of the selected event, such as the event type and time.
(2) Event detail area: displays parameter trends, compressed waveform, and parameter values of
selected event.
(3) Event summary area: displays ABD events within the
Zoom period. The selected event is enclosed in
a white frame.
(4) Event statistics area: displays the total number of ABD events and the numbers of each event within
the
Zoom period.