User's Manual

Table Of Contents
18 - 6
Red: high priority alarm event
Yellow: medium priority alarm event
Cyan: low priority alarm event
Green: manual event
White: operation-related event, such as accepting a C.O. average.
The number of currently selected events and the total number of events are displayed at the top right corner of
the event list. For example, 2/4 indicates that the selected event is the second event in the filtered events and the
total number of filtered events is 4.
Tot al indicates the total number of events. For example: Tot al : 28 means that
there are a total of 28 events.
18.2.7.2 Configuring the Filter
You can filter events by time, alarm priority, alarm category, or parameter group. To configure the filter, follow
this procedure:
1. Enter the
Events page.
Select
Filter Setup and set the desired filter criterion. Closing the Filter Setup dialog will automatically turn on
the Filter, as can be seen in the Event's tab. Events allowed by the filter will be displayed in the event list.
18.2.7.3 Editing Events
To edit events, follow this procedure:
1. Enter the
Events page and tick off the desired events.
2. Select
to edit the selected events.
Lock: manually lock the event. Locked events cannot be deleted.
Note: enter comments for the event.
Rename: allow renaming an event name. Only manual events and arrhythmia events can be renamed
if enabled by the hospital’s settings.
18.2.7.4 Viewing Event Details
To view waveforms and parameter values at the event time, follow this procedure:
1. Enter the
Events page.
2. Select
Detail.
To display beat labels on the first ECG waveform, switch on
Beat Anno. The white beat labels indicate heart
beats classification and may explain suspected, missed, or false arrhythmia calls. Heart beats are classified as
follows:
N = Normal
V = Ventricular ectopic
S = Supraventricular premature
P = Paced
L = Learning
? = Insufficient information to classify beat
I = Inoperative (for example, Lead Off)
M = Missed beat
If you switch on
Beat Anno on the Events page, beat labels will also be displayed on the Full Disclosure page,
and vise versa. Beat labels can be printed out.
18.2.7.5 Printing Event Reports
To print event reports via a printer, follow this procedure:
1. Enter the events review page.
2. Select to enter the
Print Setup dialog.